Page 13 - Shimadzu AA-7800 Series
P. 13

Any User







 Safety Technology for Flame AA Systems  Simple and Easy-to-Use Furnace System




 ˙ Automatic Flame Extinguishing Function by Vibration Detection Sensor  Replacing the Graphite Tube is Easy
 A built-in sensor automatically extinguishes the flame when it detects vibration. That eliminates any worry in the event of an   Furnace measurements require replacing the graphite
 earthquake or other major shaking.  tube. Thanks to the simple furnace structure, even
               first-time users can replace the tube easily using a
               specialized positioning jig. It is also easy to differenti-
               ate between using different graphite tubes ideal for
               given measurement applications.


                                                     Graphite Tube Positioning Jig  Replacing the Graphite Tube

 The device detects vibrations and automatically digests flames.
 Automatic flame extinguishing  Selecting the Graphite Tube
 ˙ Multi-Mode Automatic Gas Leak Check Function  Pyro-coated graphite tube  High-density graphite tube  Platform tube
                        (P/N  206-50588-11)            (P/N  206-50587-12)           (P/N  206-50887-02)
 If the power supply is ON and the flame is extinguished, this function automatically checks for gas leakage from fuel gas lines
                         Effective for elements that readily form  Effective for low-boiling point elements  Restricts chemical interference due to
 inside the system’s gas control unit. If a gas leak occurs, an alarm sound is emitted and a warning is displayed in the window.  carbides (Ni, Fe, Cu, Ca, Ti, Si, V, Mo, etc.).  (Cd, Pb, Na, K, Zn, Mg, etc.).  coexisting substances. Effective for the
                                                                                     analysis of environmental samples and
                                                                                     biosamples, such as sea water and
                                                                                     industrial waste.
 ˙ Automatic Flame Ignition and Extinguishing    * Depending on the state of the sample, some other combinations may be appropriate.
 The flame can be ignited or extinguished easily. Flashback is prevented by an Air-C2H2 flame priority ignition system.
 ˙ Automatic Air-N2O Switching Mechanism with Acetylene Flowrate Monitor
 Once the Air-C2H2 flame is ignited, it automatically switches to an N2O-C2H2 flame. If C2H2 levels do not increase, due to a   GFA-TV Graphite Furnace Camera (Optional)
 solenoid valve problem, for example, then flashback is prevented by not switching to the support gas.
               The camera can be used to confirm the sample injection position or sample drying status, which is helpful when evaluating or
               optimizing the temperature program.  The sharp image clearly shows the status inside the graphite tube.
 ˙ Acetylene Regulator Failure Detection Mechanism
 Even if the acetylene gas cylinder regulator fails, this mechanism detects the abnormally high gas pressure to protect the
 system by preventing the acetylene inlet solenoid valve from opening.

 ˙ Flashback Prevention by Pressure Monitor   ˙ Automatic Gas Shutoff by Flame Monitor

 ˙ Instantaneous Power Interruption Detection  ˙ Burner Misuse Prevention Mechanism
   Mechanism and Safety System for Reigniting Flame

 ˙ Drain Tank Level Monitor



 ˙ Flame Retardant Materials Used  ˙ Ignition Switches Designed for Safety  GFA-TV  Checking the Sample Injection Position  Checking the Drying Status
 Flame retardant materials (UL standard: 94V-0) are   The switches are shaped to fit fingers and are
 used for outer cover of the main unit and the   located near the window. The mechanism is
 atomizer. Outstanding design achieved while   designed so that ignition requires pressing
 ensuring safety.  two switches simultaneously, which eliminates
 the worry of accidental operation.  Safety Technology for Furnace AA Systems
              ɾ Cooling Water Flow Rate Monitor      ɾ Argon Gas Pressure Monitor      ɾ Furnace Block Cooling Check
 ˙ Tubing Parts with High Durability  ˙ Clear Window  ɾ Overcurrent Preventive Mechanism (Double-Checked by Circuit Protector and Optical Sensor)
  and Joints with High Reliability Used  The flame status can be checked via a large
 window and the large opening provides easy
 Tubing parts selected based on durability are
 accessibility to the burner unit. Also, the window
 connected using joints with high reliability.
 closes automatically when released, so there is
 no risk of forgetting to close the window.


 12                                                                                                               13
   8   9   10   11   12   13   14   15   16   17   18