Page 17 - Shimadzu AA-7800 Series
P. 17

Optional Accessories  Installation Conditions


               For more details, refer to the Pre-Installation Requirements.


 ASC-7800 Autosampler             Choose from 100, 120, 220, or 230 VAC,  Example of recommended piping for the atomic absorption system
 (P/N 208-00400-XX)  Power  Main unit  230 VA, 50/60 Hz
 Power requirements: 100 to 240 VAC, 50 VA, 50/60 Hz  requirements  200, 220, 230, or 240 VAC ±5%,  Stainless steel piping
                                                                           (inner diameter greater than 7 mm)
 Compact-size autosampler for flame and furnace units can install 60 samples and   GFA-7800  7400 VA, 50/60 Hz
 8 reagents.                                                                             Setting pressure
       ASC stand kit (208-00490-41) is required to mount with dedicated flame   Operating  Temperature range 10 to 35 °C  Stop cock  Stop cock   0.11 MPa
 measurement system, and ASK-7800 extension unit for furnace measurement is   environment  20 to 80% (less than 70% when
 For furnace measurement
 For flame measurement
 required to mount with dedicated furnace measurement system or dual system.  For flame measurement  For furnace measurement  Humidity range  temperature is higher than 30 °C)
       An ASC starter kit for aqueous solutions (206-59765-41) or organic solvents   1500 mm min. (W) × 700 mm min. (D)
 (206-59765-42) is required for the flame continuous method or hydride generation   Testing bench  Pressure regulator
 method.                          Withstand load: 200 kg  *1, 2  Pressure range 0.3 MPa     Acetylene gas
                       Material   Stainless  *3            Setting pressure 0.09 MPa
               Gas tube
 HVG-100 Hydride Vapor Generator  AMF-100 Atomic Absorption   Dimensions  7 mm min. I.D. × 80 mm min. (L)  Setting pressure
                                                                                            0.4 MPa
 (P/N 208-00900-XX)  Muffle Furnace (Electronic Cell Heater)  Acetylene  Cylinder: 0.11 MPa
 Power requirements: 120, 230 VAC, 50 VA,   (P/N 208-00950-47)  Lab: 0.09 MPa
 50/60 Hz  Power requirements: 200 to 240 VAC, 550 VA,   Gas supply  Compressor: 0.4 MPa
 This is useful for high-sensitivity analysis of   50/60Hz  pressure  Air  Lab: 0.35 MPa       Setting pressure
 elements such as As, Se, and Sb.  This dedicated furnace permits higher sensitivity   (set secondary  Cylinder: 0.4 MPa  Dinitrogen   0.4 MPa
       Used with an ASC-7800 autosampler, it permits   measurements using the hydride vapor generator   pressure)  Dinitrogen oxide  Lab: 0.35 MPa  Pressure range 1 MPa  oxide gas
 the automated serial analysis of up to 60 samples.  than the flame heating method. The temperature   This product contains Refractory   Setting pressure 0.35 MPa
       Nozzle ASSY, HVG (P/N: 206-67563) is required   controller provides optimal control of the quartz   Ceramic Fiber (RCF) and cannot be   Argon  Cylinder: 0.4 MPa
 for operation in conjunction with the ASC-7800.  cell temperature to prevent damages to cells due to   sold in some areas. Please inquire   Lab: 0.35 MPa
                                                             Pressure range 1 MPa
 excessive overheating.  about the details.  Material  Stainless  Setting pressure 0.35 MPa       Argon gas
 Mount adopter is required separately.
 MVU-100 Mercury Vaporizer  For AA-7800F: 206-52135-41  Dimensions  For flame: Approx. 500 mm W × 500 mm D  Setting pressure
                                                                                                    0.4 MPa
 (P/N 208-00850-XX)  For AA-7800G: 206-83755-91  Exhaust duct  For furnace: ø150 mm to ø200 mm  Mist separator
 Power requirements: 100 to 240 VAC, 50 VA,   * The AMF-100 cannot be used with the AA-7800F/AAC.  For flame: 600 to 1200 m /hr  Pressure range 1 MPa
                                               3
 50/60 Hz              Intake capacity  For furnace: 10 to 180 m 3 /hr  Setting pressure 0.35 MPa
 This mercury vaporizer is for cold vapor atomic   Atom Booster  Cooling water circulation unit or tap-water   Lab  Outdoor
 absorption spectroscopy.  (P/N 206-50957-91)  Cooling water (GFA-7800)  equipment  *4         Air compressor
       It also requires a gas flow cell (201-98687), gas   Quartz slotted tube, useful for further increasing   80 m max.
 flow cell holder (206-77703-91), and mercury   sensitivity of flame analysis with a cell holder.  *1  Atomic absorption spectrophotometer stand recommended
 hollow cathode lamp (200-38422-28).  * Cannot be used with N2O-C2H2 flame.  *2  Maintain a free maintenance space of 150 to 200 mm to the sides and rear of
 Contact us for information about the types of   the instrument.
 samples for which this option is available and   *3  Do not use pipes containing copper, silver, gold, mercury (or alloys containing
 Burner Head with Angle-scale   the elements for which this option is useful.  these metals) as pipes for acetylene.
 Markings      *4  If tap-water equipment is used, ensure that it meets the specifications below.
 (P/N 206-50370-92)  Graphite Furnace Camera GFA-TV  Compatible faucet  Faucet 13 or 12 to 15 mm O.D. rimmed faucet
 Used to specify a constant fixed angle for   (P/N 206-52950-41)  Water temperature  10 to 30 °C
 measuring high concentration elements by   Provides viewing inside the graphite tube.  Water flow rate  0.6 to 1.5 L/min
 changing the burner angle.  Including Video View Software (CD-ROM)  Supply pressure  0.08 to 0.15 MPa  *5
               Position of the supply port  Within 7 m of the instrument
 Micro Sampling Kit
 High-temperature Burner Head  (P/N 206-77540-91)  *5  If the supply pressure exceeds 0.17 MPa, use the optional Regulator ASSY.
 (P/N 206-77530-91)
 Required to use the flame micro sampling method. ASC-7800 and
 Made of pure titanium. Air-cooled. 5 cm slot for   ASK-7800 (or ASC stand kit) are also required.
 N2O–C2H2 flame
 • Extremely corrosion-resistant  O-Ring Set
 • Can also be used for Air–C2H2 flame.
 Required for analyzing organic solvents in flame analysis. Contact us for available
 organic solvents.  Dimensions
 Fluoro Rubber O-Ring Set (P/N 206-77620-92)
 Silicon Rubber O-Ring Set (P/N 206-77620-93)
                                  Wall
 Compressor and Gas Equipment
                                  Table
 Part Name  P/N  Remarks                          150 to 200
 Low-noise air compressor  208-91750-36  220/230 VAC, 50/60 Hz, with mist separator
 Mist separator kit  206-52458-41  Required if using an air compressor other than above.  AA-7800
 YR-71 compressed gas regulator  040-72020-01  For acetylene cylinder
                                                                 702  GFA-7800
 NP2-3-10B6-2RFH85-V compressed gas regulator  040-72034-01  For dinitrogen oxide cylinder (freeze-proof model)
                                                                      ASC-7800
 FR-2S-OP compressed gas regulator  208-91763  For argon cylinder  610
 Pressure regulator set, Air  208-91756-91  Pressure regulator for use in laboratory (for air)  390
 Pressure regulator set, C2H2  208-91756-92  Pressure regulator for use in laboratory (for acetylene)  870
 Pressure regulator set, N2O  208-91756-93  Pressure regulator for use in laboratory (for dinitrogen oxide)
 Pressure regulator set, AR  208-91756-94  Pressure regulator for use in laboratory (for argon)
                                                                                    340
 Hose ASSY  206-50389-41  For air used to promote incineration in GFA
                     280                                                               940
 Cooling Water Equipment                                                                         Unit: mm
 Part Name  P/N  Remarks
 CA-1116A cooling water circulator  044-01813-51  For cooling GFA, 100 VAC, 1100 VA, 50/60 Hz
 Cooler attach kit  206-84373-41  For connecting GFA and cooling water circulator
 Cooling water tube ASSY  206-51028-91  Connecting tubes when using tap water to cool GFA
 Regulator ASSY  206-86147-41  Decompression valve when using tap water to cool GFA
 16                                                                                                               17