Page 63 - LC-SFC_Pharma_Brochure
P. 63

mV
 mAU  mV                50                                       3-2.  Detailed Peak Detection Settings
 2.0
 Main peak  Impurity                                             peak integration range, and peak-baseline type may not be enough to
 [Detection Threshold] = 10  4  40                               For complex chromatograms, adjustments of the detection threshold,
 1.5  3                 30                                       obtain the desired peak integration results. i-PeakFinder is compatible
                        20                                       with a wide variety of chromatograms and allows the user to configure
 2
 1.0                                                             more detailed peak detection conditions. Some of these detailed settings
                        10                                       are described below:
 1
 0.5                    0                                        (1) To Detect Peaks Not Affected by Noise [Minimum Half Width]
 0                        1.00  1.05  1.10   min                 Smoothing  is  sometimes  performed  on  chromatograms  obtained  via
                                                                 LCMS. If the noise frequency is close to the peak frequency, then it be-
 0.0                 Fig. 5  Chromatogram with Peak Tailing (N=5)
 12.5  15.0  17.5  min                                           comes difficult to determine peaks automatically and single peaks may
 17.5  20.0  min  [Baseline Type] Not conÿgured  Table 2  Comparison of Area Reproducibility Between the Traditional  be recognized as multiple peaks. In this situation, configuring the mini-
 Normal shoulder peaks can be detected with no special parameter adjustment  mV  mum half width setting ignores noise smaller than that value and en-
                    Method and i-PeakFinder                      sures that peaks with a FWHM value above the set minimum value are
 mAU  4
 2.0                           i-PeakFinder   Traditional method  detected among the peaks with a wide wave profile due to smoothing.
 [Detection Threshold] = 10  Main peak  Impurity  %RSD  0.106  0.275  Fig. 8 shows the difference made by increasing the minimum FWHM
 [Detection Threshold of   3                                     value. This feature is useful when noise can be observed in a peak.
 1.5  Unseparated Small Peak] = 0.3  Processing  Peak-baseline Type: Base to Base  Width: 1 s
 2                        Peak-Baseline Height* 2  Slope: 2000 uV/min
                                      * Described in more detail later in the article
 1.0  1
            3. Setting Parameters
            3.  Setting Parameters
 0
 0.5
            i-PeakFinder not only performs highly accurate peak integration with de-
 12.5  15.0  17.5  min                                                   14.00  14.50  15.00  14.00  14.50  15.00
 0.0        fault settings but also includes more detailed settings for adjusting the peak
 [Baseline Type] Vertical Division                                  [Minimum Half Width] Not conÿgured  [Minimum Half Width] = 2
            detection conditions (peak width and peak-baseline start and end heights)
 mV
 17.5  20.0  min  and  peak-unifying  conditions  that  determine  how  to  combine  multiple
 Even very small shoulder peaks can be detected with additional parameter adjustments.  4  peaks. The most important of these settings will be explained here.
 Main peak  Impurity
 Fig. 2  Examples of Shoulder Peak Detection  3  3-1.  Basic Peak Detection Parameters
 2
 2-2.  Simple Adjustment of Peak-baseline   The basic parameters that allow the user to adjust peak detection condi-  14.00  14.50  15.00  14.00  14.50  15.00
 Processing  1  tions  are  the  peak-baseline  type  mentioned  above,  peak  detection   [Minimum Half Width] = 5  [Minimum Half Width] = 10
            threshold, and peak integration range. Using the detection threshold set-
            ting, the peaks below a certain threshold are not detected based on the   Fig. 8  Example Con guration of Minimum FWHM
 Accurate  detection  of  impurity  peaks  is  essential  for  pharmaceutical   0  estimated noise level calculated using a proprietary algorithm. Decreas-  (2) To Ensure Peak Area Accuracy and Linearity [Peak Baseline Height]
 quality control and other applications. Impurity peaks are often fused at   ing the peak detection threshold value allows the detection of smaller
 the base of a major peak, and quantitative results obtained via area nor-  12.5  15.0  17.5  min  peaks. The peak integration range specifies the time range during which   With tailing peaks and chromatograms with a large amount of baseline
                                                                 noise, the peak start and end points can vary depending on the data,
 malization can vary depending on the method used for peak-baseline   [Baseline Type] Base to Base  peaks will be detected. Fig. 6 shows the change in results produced by   which can reduce area accuracy. By using the peak-baseline height set-
 processing. The method used for peak-baseline processing also differs   Note: Other than peak integration range and peak-baseline processing type, all settings are at default.  changing the peak detection threshold from the default setting of 5 to   ting, the peak start and end points are recognized as the product of a
 based on the samples and testing objectives. With traditional methods,   2,000. These intuitive controls allow the user to detect or not to detect   value entered for peak-baseline height and noise intensity calculated
 performing a speciÿc type of peak-baseline processing requires the user   Fig. 4  Peak-baseline Processing Example  small peaks with a simple adjustment. Fig. 7 shows an example of adjust-  using a proprietary algorithm. Consequently, the larger the peak-base-
 to include time programming or perform peak integration manually.  ing the peak integration range. Without adjusting the peak integration   line height value is, the higher the peak-baseline is positioned.
 Table 1  Quantitative Results Obtained via Area Normalization with  range, all peaks fall within the range and the peak-baseline is affected by
 However, i-PeakFinder has adjustable parameters in its basic settings   Different Methods of Peak-baseline Type  negative peaks. Setting the peak integration range to exclude negative   Using this setting provides good reproducibility in the determination of
 for performing a speciÿc  type of peak-baseline processing, allowing   peaks allows the user to configure an appropriate baseline.  peak- baseline start and end points. This improves peak area reproducibil-
 the user to easily implement the optimum peak-baseline type in each   Not con gured  Vertical Division  Base to Base  ity, as mentioned earlier in Table 2, and provides linearity across results
 situation.  Fig.  3  shows  a  list  of  peak-baseline  type  in  the  settings   Main peak  99.681  99.448  99.680  30 mAU  30 mAU  obtained  for  a  target  component  at  different  concentrations.  Fig.  9
                                                                 shows an example configuration of peak-baseline height where the base-
 window, Fig. 4 shows the result of peak-baseline type setting of an   Impurity  0.160  0.338  0.160  25  25  line length of the tailing peaks is adjusted by specifying the peak-baseline
 impurity peak that is fused at the base of a major peak, and Table 1   20  20  height.
 summarizes the quantitative results obtained via area normalization   2-3.  Improved Peak-Baseline Positioning  15  15
 with different peak-baseline type settings. Peak-baseline processing   for Enhanced Reproducibility  10  10
 appropriate  for  a  particular  situation  can  be  simply  performed  by   5  5
 changing a few basic parameters.  0   0
 When a target component is separated with peak tailing or leading in   1.0  2.0  min  1.0  2.0  min
 some cases, using a traditional peak integration method would result   [Detection Threshold] = 5 (default value)  [Detection Threshold] = 2000
 in variability in the calculated peak areas depending on which point   Fig. 6  Peak Detection Threshold
 along the trailing and leading edge was deÿned  as the peak base.         15.0      min     15.0      min
 This affected the reproducibility of area results.                           [Peak Baseline Height] Not conÿgured
 With i-PeakFinder, the user can adjust the height at which the start
 and end of a peak-baseline needs to be deÿned so that peak-baseline
 processing is consistent even with tailing and leading peaks.
 The reproducibility of results obtained using the traditional method and
 i- PeakFinder is summarized in Table 2 for the chromatogram shown in Fig.
                                                     min
                                                10.0
                                           5.0
                    5.0
                         10.0
                              min
 5. In the example below, the peak area results obtained using i-PeakFinder   [Peak Integration Range] Not conÿgured  [Peak Integration Range] ° 4.1 (min)  15.0  min  15.0  min
 exhibit superior reproducibility compared to the traditional method.            [Peak Baseline Height] = 50
 Fig. 3  Peak-baseline Type Settings  Fig. 7  Peak Integration Range
 2                                                               Fig. 9  Example Con guration of Peak-Baseline Height for Tailing Peaks  3
   58   59   60   61   62   63   64   65   66   67   68