Page 54 - Clinical-Medical Solution for Clinical Chemistry
P. 54

Integration of amino acid, acylcarnitine and steroids
       analysis in single FIA/LC-MS/MS platform






                  Conc.                         LOD    LOQ               Conc.                        LOD    LOQ
                  (ng/mL)  Area  %RSD    S/N   (ng/mL)  (ng/mL)         (ng/mL)  Area   %RSD   S/N   (ng/mL)  (ng/mL)
                    5     15,919  6.05   66.1   0.23   0.76               5     97,910  0.62  184.1   0.08   0.27
                    25    54,064  2.11  246.3                             25    387,155  0.92  946.5
                    75   186,970  0.85  681.0                             75   1,247,236  0.08  2338.6
           Cortisol                                               4-AD
                   125   322,782  1.68  971.9                            125   2,187,611  0.26  5318.0
                   150   345,172  0.26  994.9                            150   2,406,566  0.36  4459.4
                   500   1,307,317  0.16  1606.4                         500   9,143,414  0.69  36564.1

                    5     23,687  1.87   27.8   0.54   1.80               5     40,107  6.12   79.3   0.19   0.63

                    25    98,627  3.64  103.0                             25    157,773  2.85  366.5
                    75   388,221  2.04  473.1                             75    504,027  1.03  830.9
           21-DOF                                                17-OHP
                   125   701,690  1.16  181.9                            125    874,209  1.56  3824.8
                   150   758,860  0.43  537.1                            150    951,763  0.62  1635.5
                   500   3,000,289  0.70  214.5                          500   3,694,638  0.20  1932.9

                    5     39,161  1.53   60.5   0.25   0.83

                    25   167,525  2.43  218.1
                    75   534,404  0.39  436.3
           11-DOF
                   125   964,525  0.26   94.2
                   150   1,029,424  0.53  695.0
                   500   3,872,397  0.60  97.9






          Conclusions


          This platform is an effective tool for an initial screening and able to minimize sample consumption down to 1 uL.
          However, the use of a high performance LC/MS/MS system is highly recommended in order to achieve the appropriate
          level of sensitivity for the steroids especially. Using high performance LC/MS/MS (LCMS-8050, Shimadzu Corporation,
          Japan), the observed limit of detection (LOD) for the analysis of 17-hydroxyprogesterone was 0.19 ng/mL.














                                                                                                      First Edition: May, 2015


                                               For Research Use Only. Not for use in diagnostic procedures.
                                               The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
                                               The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
                                               accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
                                               use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
                                               to change without notice.
       www.shimadzu.com/an/                                                                        © Shimadzu Corporation, 2015
   49   50   51   52   53   54   55   56   57   58   59