Page 52 - Clinical-Medical Solution for Clinical Chemistry
P. 52

Integration of amino acid, acylcarnitine and steroids
       analysis in single FIA/LC-MS/MS platform






          DBS samples (d = 5mm) were placed in 96-well plates,   ultrasonic bath for 30 min. Samples were measured using
          and AAs, ACs and steroids were extracted with 180  L of   a Nexera UHPLC system coupled to LCMS-8050 triple
          80% acetonitrile-water solution consists of the known   quadrupole mass spectrometer (Shimadzu Corporation,
          concentrations of stable isotope labeled standards of each   Japan).
          compounds. The extraction were performed in an




          Result


          DBS provides a number of advantages, for examples, a   using the combination of Nexera MP and LCMS-8040
          less invasive and much simpler sample collection      (Shimadzu Corporation, Japan). In addition to that, we
          method rather than venipuncture technique.            independently developed a method for steroids in DBS.
          Furthermore, it provides you simpler storage and      Steroids were separated on a Phenomenex kinetex
          transportation as well as it can lower the infection risk   XB-C18 (50x2mm, 2.6 m) at a column temperature of
          of various pathogens, and requires a smaller blood    40 °C for 5 min.
          volume. To date, DBS-LC-MS/MS has emerged as an       In this study, we present a strategy for performing AAs,
          important method for quantitative analysis of small   ACs and steroids analysis within a single LC/MS/MS
          molecules.                                            platform. AAs, ACs and steroids were extracted from
          Previously we developed an innovative AAs and ACs     only one dried blood spot. This system enables to
          screening method makes it possible to inject just 1 µl   automatically analyze 7 min in all target analytes in 2
          of sample and successfully reduce analytical run time as   injections.
          fast as 74 seconds (conventional method >120 sec.)

                         B conc. (%)                                                  Flow rate (mL/min)
                         100    90                                                  90         1.0
                                                               Injection
                                        0.7                    for steroids
                                              50         50           55
                          50                                                                   0.5
                             Injection                   0.3
                             for AAs, ACs
                             0.13
                           0                                                                   0
                            0       0.5     1.0      1.5   0   1.0   2.0   3.0   4.0   5.0   6.0   7.0
                                  1 min               min                   4 min             min
                               Analysis of     wash      equilibration    Analysis of steroids     wash
                               AAs and ACs
                                                     valve                                    valve
                                                   switching                                switching







                                     Figure 2  Time Program of amino acids, acylcarnitines and steroids analysis









                                                                                                                      5
   47   48   49   50   51   52   53   54   55   56   57