Page 7 - Solutions for Plastic Evaluation
P. 7

Structure and Composition                                   Evaluation of Raw Materials

                                                                                                           FTIR



               FTIR Measurement of Plastic Pellets and Powder by Single Reflection ATR Method
               The figure on the left below shows results from measuring various types of plastic pellets and powders. The single reflection ATR method
               allows measuring pellet and powder samples directly by simply placing them tightly against an approximately 2 mm diameter prism.   Plastic Materials  Evaluation of
               Furthermore, it allows even new users to easily obtain qualitative information about measured polymers by using the spectral search function.



                     ABS


                     Polycarbonate                                                                                    Raw Materials  Evaluation of



                     PET

                     Nylon 6



                     Polyphenylene sulfide                                                                             Product Evaluation

                                                                               Spectral Search Window
                       Example of Measuring Plastic Pellets and Powders  Example of Measuring Plastic Pellets and Powders




               Single Reflectance ATR Method                                                                           Product Information
               Single reflection ATR measures the infrared spectrum of a sample pressed against the surface of a prism approximately 2 mm in diameter.
               It allows measuring samples by simply placing them in close contact with the center of the prism. Therefore, it is not restricted to flat
               samples and can measure curved or powdered samples as well. Liquid samples can be measured by simply dripping them onto the prism.
               A photograph of the DuraSamplIR II single reflection ATR attachment and an illustration of its optics are shown below.  It irradiates the
               sample with infrared light through the prism and obtains information about the sample surface (to a depth of about 1 µm) based on the
               light reflected from the boundary surface. Not only does this eliminate the need for dilution or other sample pretreatment, but also allows
               obtaining information about the sample surface easily.






















                             Photograph of DuraSamplIR II                       Diagram of Optics




                                                                                                   LAAN-D-XX005       7
   2   3   4   5   6   7   8   9   10   11   12