Evaluation of Plastic Materials
                             Quality Control
              FTIR



    Evaluation of   Plastic Materials  Determination of Hydroxyl Value in Polypropylene Glycol by NIR-PLS Method

               Polypropylene glycol (PPG) is a liquid manufactured by polymerizing propylene glycol and is used as a raw ingredient of polyurethane.
               The hydroxyl value is an index of the content of free hydroxyl groups in polyols, such as PPG. JIS K 1557 (Plastics Polyols for use in the
               production of polyurethane) specifies using titration to measure the hydroxyl value. However, titration requires equipment and reagents,
               and is also time-consuming. In contrast, the NIR-PLS (near infrared-partial least squares) method allows calculating the hydroxyl value easily
               and quickly by simply measuring the near infrared spectrum of the sample, then using PLS to calculate quantities.

    Evaluation of   Raw Materials                H ʢOCH2CHʣ n - OH


                                                               -
                                                 ɹɹɹɹ CH3
                                                  Structural Formula of Polypropylene Glycol


               NIR spectra for four homologous types of PPG with different hydroxyl values and a calibration curve created using the PLS
     Product Evaluation  method are shown below. The NIR spectra indicate a difference in their peak intensities near 6990 cm  and 4850 cm .  The
                                                                                                        -1
                                                                                             -1
               peaks near 6990 cm  correspond to the first overtone of the O-H stretching vibration, whereas the peaks near 4850 cm
                                                                                                            -1
                              -1
               correspond to a combination of the O-H stretching and bending vibrations.

                          1.75
                           Abs
                           1.5
     Product Information  1.25 1                                                      Corr. Coef.: 0.99996



                          0.75
                                                                                      MSEP         : 0.00008

                                               6000  5500  5000 4500                  SEP     : 0.00901
                                                      1/cm
                               NIR Spectra of Polypropylene Glycol   Actual Hydroxyl Value (known concentration) vs.
                                                                      Predicted Hydroxyl Value (quantitative value)



               NIR Heating Transmission Cell

               This cell allows measuring transmittance while liquid samples placed in the provided
               6-mm  diameter  test  tube  are  heated  or  held  at  a  constant  temperature.  This  is
               especially  convenient  for  quantitative  analysis  of  multiple  components  in  liquid
               samples or when monitoring reactions during heating processes.


               Specifications
               • Measurement range     : 12,500 to 3,800 cm -1
               • Temperature range     : Room temp. to 120°C
               • Accessory recognition function : Included
               • Test tube               6 mm dia. × 50 mm