Page 29 - Oligonucleotide Therapeutics Solution Guide
P. 29

Characteristic analysis
 Quality Control

 Quantitation & Spectrum Confirmation  BioSpec-nano




                 Features
 Quantitation of Nucleic Acid Using
               The BioSpec-nano is ideal for measuring precious biological samples, such as nucleic acids and proteins, as it can measure as little as 1 ~ 2 μL   Modification
 BioSpec-nano  without using a cell (cuvette).                                                                     Target selection
               The measurement is started simply by dropping the sample into the dropping position (target) and pressing the start button. In addition,
 click here    the automatic wiping function does not require sample wiping after measurement. While maintaining high correlation with standard type
               spectrophotometer (Cell Measurement, Double Beam), high reproducibility and measurement accuracy have been achieved. The optical path length
               can be selected from 0.2 mm and 0.7 mm depending on the sample concentration. In addition, if the option is purchased, measurement using a cell
 •    It can be used to measure the concentration of extracted double-stranded   with an optical path length of 5 mm (sample volume of 2 mL) is also possible.
 DNA and to check the degree of purification.  The BioSpec-nano is equipped with a variety of quantification modes, including nucleic acid quantification, protein quantification, OD value display
               at specified wavelengths, and can be used in a variety of applications.
 benefits  •    It is possible to measure from a minimum sample capacity of 1 µL.
 •    It is equipped with an automatic wiping function that achieves low carryover.                              Unprotected  Excision


  Methods and Results                                                                                                Oligomer synthesis

 Sample  Double-stranded DNA
 Conc., Volume As shown in Table 1
 Preparation  Samples were prepared using Tris-EDTA (TE) buffers at the
 concentrations shown in Table 1.
 Analytical   Samples were measured to 10 times of optical density (OD) at
 Conditions  260 nm at each optical path length (Table 1).
 The result of   The correlation coefficient between the measured value and the   You can measure samples one after another while
 optical path   reference value was 0.999 (Figure 1). When the concentration   holding the micropipette in your hand.
 length 0.2 mm was 250 ng/μL or more, the reproducibility of measurements was
 1.4% or less for CV values (%) and -5.4% to 2.8% for OD errors                                                  Purification
 (%).
 The result of   The correlation coefficient between the measured value and the
 optical path   reference value was 0.999 (Figure 2). When the concentration   Figure 1   The result of optical path length 0.2 mm
 length 0.7 mm was 70 ng/μL or more, the reproducibility of measurements was
 1.4% or less for CV values (%) and -8.6% to 4.4% for OD errors
 (%).          Automatic wiping function prevents
 The result of a  The correlation coefficient between the measured value and the
 5 mm optical   reference value was 0.999 (Fig. 3). When the concentration was   forgetting to wipe
 path length cell 70 ng/μL or more, the reproducibility of measurements was 0.6%
 or less for CV values (%) and -1.6% to 3.6% for OD errors (%).  We measured 3 μL of double-stranded DNA with an optical path
               length of 0.7 mm and 578 ng/μL, followed by one wiping operation,
 Table 1  Analysis Conditions
               and then measured TE buffer to determine carryover (%) (Equation 1).
 Concentration  50 ~ 3700 ng/µL  The wiping operation is set to be performed once after the sample or
 Opticall path length 0.2 mm
 Volume  1 µL  TE buffer measurement. As a result of performing 60 sets of double-                               Quality Control
 Concentration  15 ~ 1000 ng/µL  stranded DNA measurement → wiping → TE buffer → wiping in 1 set,
 Opticall path length  0.7 mm                                                                                      Characteristic analysis
 Volume  2 µL  the carryover (%) was kept at 0.3% or less, and it was confirmed that
 Concentration  2 ~ 150 ng/µL  the remaining sample amount was very low when automatic wiping
 5 mm pathlength cell (Option)  Figure 2   The result of optical path length 0.7 mm  The device automatically sets / measures / wipes the
 Volume  2 mL  was used.
                                                                   optical path length.
               Carryover (%)
                          [Nucleic acid concentration in TE measurement]
               = 100 ×                                                  · · · Eq.1                               DDS
                        [Nucleic acid concentration in dsDNA measurement]                                          Pharmacokinetics




               Various quantitative modes applicable to various samples

 Figure 3   The result of a 5 mm optical path length cell
               Quantitative mode  :  This mode is used to quantify RNA, dsDNA, ssDNA, OligoDNA, etc. and calculate nucleic acid concentration,
                                 OD ratio (OD 260/280, OD 260/230).
               Labeled nucleic acid   :  This mode is used to quantify nucleic acids labeled with Cy3 and others (RNA, dsDNA, ssDNA, OligoDNA) to
               quantification    calculate nucleic acid concentration, label concentration, nucleotide concentration, and labeling rate.
  Conclusions                                                                                                    Other
               Protein quantification  :  This mode finds the protein concentration from the absorbance at 280 nm. In addition, it is possible to
                                 determine the label concentration and labeling rate for labeled proteins as well as nucleic acids.
 Bipspec-nano enables nucleic acid quantification with excellent photometric linearity, reproducibility, and accuracy at optical path lengths of 0.2 mm
 and 0.7 mm with sample volumes of 1 ~ 2 μL, respectively.  OD display   :  This mode displays the OD value for any specified wavelength, up to a maximum of 8 wavelengths.
               for specified wavelength


 28                                                                                                               29
 index                                                                                                   index
   24   25   26   27   28   29   30   31   32   33   34