High-Performance Autosamplers

            Facilitate Ultra High Sensitivity Analysis



                                                                          pr
                                                                             ov
                                                                               ed
                                                                        Im
                Popular Low-Carryover Performance Now Further Improved
                                                               Further

                                                                               ed
            Nexera X2 series autosamplers are carefully designed to reduce carryover in increasingly y
                                                              gl
            high-sensitivity analytical applications, with features such as a multi-rinse mechanism,
            included as standard, that allows you to select from multiple types of solvents. They also
                                                               so
            offer high accuracy at the injection volumes optimal for column sizes typically used in
                                                              exera
            UHPLC applications. Together with the broad injection volume linearity, this makes Nexera
            X2 series autosamplers more than capable of handling applications from UHPLC to HPLC.
                                                               LC.
                Multi-Rinse Mechanism Eliminates Carryover
            When analyzing multiple components simultaneously, they often have significantly different polarities. In such cases, a single rinse
            solution is insufficient for adequate rinsing. To solve this problem, Nexera X2 series SIL-30AC and SIL-30ACMP autosamplers feature a
            hardware structure that resists adsorption as well as a rinse mechanism that minimizes carryover.
                 Multi-rinse mechanism supports up to 4 solutions
                 The needle surface can be rinsed by a maximum of two rinse solutions   Sample loop
                 and the needle inner surface can be rinsed by a maximum of three   Needle
                 rinse solutions. Of course, rinsing with just one solution in the usual
                 way is also possible. The rinsing order can be set as desired.          Rinse ports


                        Contact surfaces of needle and needle port
                          where carryover is likely to occur            Drain
                                                                                  R0  R1  R2      R3
                                                                               For rinsing needle surface and  For rinsing needle
                                                                                 inner surface (3 types)  surface (1 type)
                                                         Needle port
                                                       (schematic diagram)          Multi-rinse mechanism can
                                                                                   accommodate up to 4 solutions
                         Port section is rinsed by any of the rinse solutions  Needle port rinsing mechanism
                                (1 to 3 solutions)
                                                                  The needle port is one of the places on an autosampler
                                                                  where carryover is likely to occur. On the SIL-30ACMP, the
                                             Discharge to drain port  needle port is rinsed automatically.

            HPLC detectors, such as mass spectrometers, are becoming increasingly sensitive each year. Consequently, the carryover levels required
            for HPLC systems are becoming extremely strict. Nexera X2 series SIL-30AC and SIL-30ACMP autosamplers offer the ultimate in
            low-carryover performance, due to the thorough carryover-reducing mechanism.

            Ultralow carryover achieved without rinsing        Ultralow carryover even on a high-sensitivity LC/MS/MS
            When rinsing is performed to keep carryover low, the total analysis time   Ultralow carryover performance is required with LC/MS systems.
            sometimes increases as the number of analyses increases. Nexera   The SIL-30ACMP demonstrates exceptional carryover performance even on
            autosamplers excel in suppressing carryover even without rinsing.  compounds such as chlorhexidine that are very prone to adsorption.
                                                               Moreover, the SIL-30ACMP features an improved rinsing mechanism to
                                                               achieve even lower carryover.
               mAU
                 5.0                                             (x10,000)
                 4.0                                               2.0
                                        0.0004 % without rinsing                      Chlorhexidine 500 ng/µL
                 3.0                                               1.5                       0.0001 %
                 2.0
                                                                   1.0
                 1.0
                                         Caffeine 4 g/L            0.5
                 0.0
                                      Blank
                                                                   0.0                       Blank
                   0    1.0   2.0    3.0   4.0   5.0    min         0.00    0.25   0.50   0.75    1.00    min


      8
   3   4   5   6   7   8   9   10   11   12   13