Page 7 - Shimadzu Nexera X2
P. 7

Not Just Sensitivity — High Data Reliability

            The SPD-M30A adopts the new temperature-controlled TC-Optics   mAU
            with the SR-Cell, which has an optimized heat exchange inlet   10                 SPD-M30A
            pipe. These features achieve faster stabilization with low external   0
            dispersion for UHPLC analysis.
                                                                       -10
            The faster stabilization shortens the “wait time” for analysis,
                                                                       -20
            thereby increasing total analytical throughput.
                                                                                            Other vendor's
                                                                       -30                  PDA detector
                                                                       -40
                                                                         0   10  20   30  40   50  min
                                                                             Baseline after instrument startup

                Excellent Stability Even with Subtle Changes in Room Temperature

            Subtle fluctuations in room temperature are unavoidable, even in   mAU
                                                                      25                         254 nm
            temperature-controlled laboratories. Incorporating TC-Optics,   20
                                                                         Room                         30°C
            which provides a stable baseline even if the room temperature   15 temperature
                                                                      10
            fluctuates, the SPD-M30A provides highly reliable data.    5                       SPD-M30A  25°C
                                                                       0
                                                                       -5
                                                                      -10
                                                                      -15                  Other vendor's
                                                                                           PDA detector
                                                                      -20
                                                                        0   100  200  300  400  500  600 min
                Superior spectrum resolution and linearity

            The SPD-M30A provides unrivaled spectrum resolution by   In addition, the new signal treatment technology realizes
            adopting an optimized optic system. A benzene spectrum is   excellent spectrum linearity across a wide area from low to high
            often used for evaluation of spectrum resolution.   concentration. The SPD-M30A supports analyses, such as purity
            The SPD-M30A offers the world’s highest* spectrum resolution,   analysis, that require a wide dynamic range.
            B/A = 2.90.
                              * As of November 2012, according to Shimadzu survey.
                                                                mAU
                                                                                    Caffeine 1000 mg/L
                            B/A = 2.90
                                                                                    Caffeine   100 mg/L


                                                B
                                                                                  Similarity 0.9998

                                             A

                          0
                              230 240 250 260 270 280 nm                250      300      350     nm
                                                                                             (normalize)


                SPD-M30A — Ideal for Both UHPLC and HPLC

            Needless to say, Nexera family SPD-M30A detectors can be installed in Shimadzu
            Prominence series HPLC systems as well. Each year, users continue to demand higher
            sensitivity, not only for food products and chemicals, but in other fields as well.
            The SPD-M30A detector is ideal for all sorts of fields.



                                                                                               SPD-M30A



                                                                                                                     7
   2   3   4   5   6   7   8   9   10   11   12