Page 14 - Shimadzu Nexera Prep
P. 14

High Separation via Preparative Recycling




                     Components Difficult to Separate can be Recovered at High Purity Levels and at Low Cost
                     Recycling Preparative System


                       What is the Recycling Separation Method?


                     Because long preparative columns are expensive, there is a need to use comparatively lower cost short columns.
                       In the recycling separation method (closed valve recycling), the eluate liquid containing the target components that has eluted
                     from the separation column is recycled into the column, enabling an equivalent separation capacity to that of a longer column.
                                  Solvent delivery unit                      Detector
                                                                Column

                                                  Injector
                                                                                        Recycle
                                                                                        valve





                                                                                              Outlet
                                                       Recycle  ow channels                    ow
                                                                                              channel
                                      Mobile phase
                                                Flow Lines for the Recycling Preparative System


                     The figure below shows the results of a seven-cycle recycling separation. In the first injection (typical separation), the separation
                     of the two components is insufficient (red area). However, when the column eluate is returned to the column from the detector,
                     it is separated a second time. If this recycling is repeated, the results obtained are equivalent to connecting a number of columns
                     in series corresponding to the number of repetitions. In this example, a 4.0 or better resolution was ultimately obtained with
                     seven recycling separation cycles (blue area).

                                        First Elution




                                              Second
                                                        Third
                                                                  Fourth
                                                                            Fifth
                                                                                      Sixth
                                                                                               Seventh





                           0.0    10.0    20.0   30.0    40.0   50.0   60.0    70.0   80.0    90.0   min

                        Example of the improvement in separation by recycling: The coeluting peaks (red) are completely separated (blue).
                                    Flowrate         : 10 mL/min
                                    Detection wavelength : 254 nm
                                    Column           : Shim-pack ™  PREP-ODS(H) 20 mm I.D. × 250 mm L.
                                    Mobile phase     : Water/methanol = 1/9 (v/v)
                                    Sample           : Mixed 1% n-butylbenzene/iso-butylbenzene solution




        14
   9   10   11   12   13   14   15   16   17   18   19