Page 7 - Shimadzu DTG-60 Series
P. 7

Thermomechanical Analyzer  TMA


 TMA-60 Series


 High precision measurement accomplished using  Accurate automatic length measurement  Theory and operation of the TMA-60
 simple operations. Even more functions.  The high precision displacement sensor can be used to measure the length
            of the sample. Compared with the other methods calculated from the
 The TMA-60 series thermomechanical analyzer accommodates a wide variety of samples   moving distance of motor, more accurate length measurement is possible.   Spring
 and is able to use various types of measurement methods* (expansion, elongation, or
 penetration) to thoroughly evaluate sample characteristics. Newly function, such as the   Force Coil
 automatic length measurement and safety features, were incorporated into   A Wide Variety of Loading Programs
 development, resulting in high performance, high functionality and ease-of-use in many   In addition to applying static loads to samples, constant rate loading,
 dimensions.  constant rate elongation or cyclic loading programs can be selected.   Displacement
                                                                                 Sensor
 * Model TMA-60 is capable of the total expansion method and  Therefore, it is able to measure stress-strain curves or thermal stresses in   Detection Probe
 * Model TMA-60H the differential expansion method.   lm or  ber samples.   Actuation Motor
             TMA                              LOAD
              um                              g
                                      3.67 g   5.00                           Micrometer
                  Shrinkage stress
           100 .00                                                           Thermocouple
                                               0.00
                                                                                Furnace
 Easy operation   Displacement of sample       -5 .00
 Changing between measurement mode is easy and maintainability is   Sample  Probe  Probe  -0 .00  Sample
 Holder
 outstanding thanks to the use of a bayonet type sample holder that can   Probe
 be attached or removed in one step and plug-in type temperature sensors.  Sample  -10 .00
 Sample             50 .00   100 .00   150 .00                      Thermo mechanical analysis is the measurement of a materials
                             Temp [゚C]
 Sample           Measurement of film shrinkage stress              behavior, expansion and/or contraction, as a function of an applied
 Chuck  Table                                                       load or temperature. A scan of dimensional changes related to time
 Elongation  Expansion  Penetration                                 or load provides invaluable information about the samples
            Options                                                 mechanical properties. The advanced design of the TMA-60
 TMA measurement mode                                               provides a wide dynamic measuring range through superior
              Low-Temperature Furnace                               integration of detection probe, displacement sensor and force coil.
              LTB-60 for TMA-60 Analyzers
              (P/N 347-65002-93)
 Probe Safety Function  High precision and Wide dynamic range  Used with a TMA-60 analyzer, it allows
 TMA-60 is programmed with a safety mechanism to prevent sample from   Measurement precision was increased dramatically by using a specialized   measurements between −150 and 600°C.
 sticking to detection probe, such as when heating glass. When   high precision digital displacement sensor. At the same time, a wide span   Liquid nitrogen gas is used below room
 displacement exceeds a set range, such as due to a sample melting, the   of ±5 mm is possible (twice as much as previous), allowing high precision   temperature.
 measurement is immediately stopped and the load is removed from the   measurements of deformations ranging from tiny to large.
 sample.
 TMA %
 618.74˚C
            Specifications
 0.60                                                TMA-60                               TMA-60H
                                            Ambient to 1000°C (Expansion Mode)         Ambient to 1500°C
             Temperature range
 468.84˚C                                     −150 to 600°C (using LTB-60)
 0.40
 557.38˚C    Measurement range                                 Displacement: ± 5 mm, Load: ± 5 N
             Sample load                                             0 - ± 5 N (500 gf)
                                                ø8 mm x 20 mm or less                 ø5 mm diameter or less
 0.20        Sample size (Film Samples)
                                         (5 mm wide x 1 mm thick x 20 mm long or less)  5 - 20 mm length
             Prove/Support tube                      Quartz                               Alumina
             Measurement mode               Expansion, Elongation, or Penetration     Differential expansion
 0.00
                                                                Constant rate Load up to 50 steps
 Flexible Parameter Settings
 0.0  200.0  400.0  600.0  Loading Mode                       Constant rate Elongation up to 50 steps
 Temp [˚C]                                                            Shrink Stress
                                                                Cyclic Load 0.01-1 Hz Frequency
             Atmosphere                                              Air and inert gas
             Dimensions and weight                             W 367 x D 624 x H 880 (mm), 45 kg
             Power supply                 AC 100 V, 120 V, 230 V  1000 VA, 50/60 Hz  AC 100 V, 120 V, 230 V  1500 VA, 50/60 Hz


                                                                                                        60 Series
 6                                                                                                 Thermal Analysis Instruments  7