Page 6 - Shimadzu DTG-60 Series
P. 6

Thermomechanical Analyzer          TMA


                                            TMA-60 Series


                                            High precision measurement accomplished using                                             Accurate automatic length measurement                        Theory and operation of the TMA-60
                                            simple operations. Even more functions.                                                   The high precision displacement sensor can be used to measure the length
                                                                                                                                      of the sample. Compared with the other methods calculated from the
                                            The TMA-60 series thermomechanical analyzer accommodates a wide variety of samples        moving distance of motor, more accurate length measurement is possible.   Spring
                                            and is able to use various types of measurement methods* (expansion, elongation, or
                                            penetration) to thoroughly evaluate sample characteristics. Newly function, such as the                                                                      Force Coil
                                            automatic length measurement and safety features, were incorporated into                  A Wide Variety of Loading Programs
                                            development, resulting in high performance, high functionality and ease-of-use in many    In addition to applying static loads to samples, constant rate loading,
                                            dimensions.                                                                               constant rate elongation or cyclic loading programs can be selected.   Displacement
                                                                                                                                                                                                           Sensor
                                            * Model TMA-60 is capable of the total expansion method and                               Therefore, it is able to measure stress-strain curves or thermal stresses in   Detection Probe
                                            * Model TMA-60H the differential expansion method.                                         lm or  ber samples.                                            Actuation Motor
                                                                                                                                       TMA                              LOAD
                                                                                                                                        um                              g
                                                                                                                                                                3.67 g   5.00                           Micrometer
                                                                                                                                            Shrinkage stress
                                                                                                                                     100 .00                                                           Thermocouple
                                                                                                                                                                         0.00
                                                                                                                                                                                                          Furnace
            Easy operation                                                                                                                  Displacement of sample       -5 .00
            Changing between measurement mode is easy and maintainability is   Sample  Probe  Probe                                   -0 .00                                                              Sample
                                                                 Holder
            outstanding thanks to the use of a bayonet type sample holder that can   Probe
            be attached or removed in one step and plug-in type temperature sensors.                       Sample                                                        -10 .00
                                                               Sample                                                                         50 .00   100 .00   150 .00                      Thermo mechanical analysis is the measurement of a materials
                                                                                                                                                       Temp [゚C]
                                                                            Sample                                                          Measurement of film shrinkage stress              behavior, expansion and/or contraction, as a function of an applied
                                                                Chuck       Table                                                                                                             load or temperature. A scan of dimensional changes related to time
                                                                     Elongation  Expansion   Penetration                                                                                      or load provides invaluable information about the samples
                                                                                                                                      Options                                                 mechanical properties. The advanced design of the TMA-60
                                                                               TMA measurement mode                                                                                           provides a wide dynamic measuring range through superior
                                                                                                                                        Low-Temperature Furnace                               integration of detection probe, displacement sensor and force coil.
                                                                                                                                        LTB-60 for TMA-60 Analyzers
                                                                                                                                        (P/N 347-65002-93)
            Probe Safety Function                              High precision and Wide dynamic range                                  Used with a TMA-60 analyzer, it allows
            TMA-60 is programmed with a safety mechanism to prevent sample from   Measurement precision was increased dramatically by using a specialized   measurements between −150 and 600°C.
            sticking to detection probe, such as when heating glass. When   high precision digital displacement sensor. At the same time, a wide span   Liquid nitrogen gas is used below room
            displacement exceeds a set range, such as due to a sample melting, the   of ±5 mm is possible (twice as much as previous), allowing high precision   temperature.
            measurement is immediately stopped and the load is removed from the   measurements of deformations ranging from tiny to large.
            sample.
                                                                  TMA %
                                                                                                       618.74˚C
                                                                                                                                      Specifications
                                                                   0.60                                                                                                        TMA-60                               TMA-60H
                                                                                                                                                                      Ambient to 1000°C (Expansion Mode)         Ambient to 1500°C
                                                                                                                                       Temperature range
                                                                                               468.84˚C                                                                 −150 to 600°C (using LTB-60)
                                                                   0.40
                                                                                                      557.38˚C                         Measurement range                                 Displacement: ± 5 mm, Load: ± 5 N
                                                                                                                                       Sample load                                             0 - ± 5 N (500 gf)
                                                                                                                                                                          ø8 mm x 20 mm or less                 ø5 mm diameter or less
                                                                   0.20                                                                Sample size (Film Samples)
                                                                                                                                                                   (5 mm wide x 1 mm thick x 20 mm long or less)  5 - 20 mm length
                                                                                                                                       Prove/Support tube                      Quartz                               Alumina
                                                                                                                                       Measurement mode               Expansion, Elongation, or Penetration     Differential expansion
                                                                   0.00
                                                                                                                                                                                          Constant rate Load up to 50 steps
                           Flexible Parameter Settings
                                                                      0.0       200.0      400.0       600.0                           Loading Mode                                     Constant rate Elongation up to 50 steps
                                                                                      Temp [˚C]                                                                                                 Shrink Stress
                                                                                                                                                                                          Cyclic Load 0.01-1 Hz Frequency
                                                                                                                                       Atmosphere                                              Air and inert gas
                                                                                                                                       Dimensions and weight                             W 367 x D 624 x H 880 (mm), 45 kg
                                                                                                                                       Power supply                 AC 100 V, 120 V, 230 V  1000 VA, 50/60 Hz  AC 100 V, 120 V, 230 V  1500 VA, 50/60 Hz


                                                                                                                                                                                                                                  60 Series
      6                                                                                                                                                                                                                      Thermal Analysis Instruments  7
   1   2   3   4   5   6   7   8   9   10   11