SCA-130-502



                                                                POC-Peak:















                                                                NPOC-Peak:




                  Fig. POC-Calibration with IC-standard


            ■ System testing
            In POC analysis, it is of great importance that
            the LiOH trap functions perfectly. A test
            should,  therefore,  be  carried  out  to  confirm
            the efficiency of the CO 2 trap.  This test
            should be carried out each working day:             Result:
            An  IC  control  solution  (TIC  =  1000  mg/L)  is
            prepared  and  analyzed  as  a  POC  sample.        TOC=4,05mg/l
            For efficient functioning of the LiOH trap, the
            POC measuring result must be < 0.1 mg/L.            POC=4,05mg/l   =>  100%
                                                                NPOC=0mg/l

            ■ Example of a POC measurement

            Sample: Toluene in ultrapure water                  ■ Recommended Analyzer / Configuration
                                                                TOC-L CXX
            Toluene  is  a  compound  that  is  fully           ASI-L
            purgeable.  In  the  addition  method,  the         POC-Option
            toluene  sample  is  completely  determined  as
            POC.


            POC measuring method:
            Injection volume:     800 µL
            Purging time:         3 min