Sum parameter – Total Organic Carbon



                                         Determination of the purgeable organic

                                         carbon (POC)
                    SCA-130-502


            According to EN 1484, which  contains the           Here,  the  volatile  organic  compounds  are
            instructions for TOC determination, the POC         converted to CO 2 and detected via  NDIR
            (purgeable organic carbon) is the TOC               detector.
            content that can be expelled under the
            conditions of this method. This information is
            very unspecific and should be described here                           TOC
            in more detail.

            The instruments in the TOC-L series can be                    NPOC                     POC
            extended    with   an   option   to   include
            measurement of  the POC parameter. The
            core  feature  of  this  option  is  the  LiOH  trap,
            which is placed in the flow line of the                 Fig. The addition method (NPOC + POC)
            analyzer.                                           During the course of the addition method, the
                                                                remainder is used for  NPOC determination.
                                                                However, in the drinking water application or
                                                                ultrapure water application, the POC content
                                                                is completely negligible. Here, TOC = NPOC

                                                                The  POC  can,  nevertheless,  play  an
                                                                important  role  in  wastewaters,  particularly  in
                                                                industrial effluents.


                                                                ■ Calibration
                                                                Particular  attention  should  be  paid  to
                                                                calibration  of the POC. Standards that are

                                                                 prepared with volatile, purgeable organic
                                                                substances are inherently very unstable. This
            For POC determination, the  sample is               is  why  IC  standard  solutions  (prepared  from
            aspirated  using  the  TOC-L  injection  syringe,   carbonates and/or hydrogen carbonates) are
            acidified  with  HCL  and  subsequently  purged     used for POC calibration. Sampling of the IC
            using carrier gas. In this step, CO 2 originating   solutions for POC determination is carried out
            from carbonates and hydrogen carbonates as          using the  TOC-L injection syringe. The IC
            well as all volatile organic compounds (POC)        solution  is  acidified  in  the  syringe.  The
            are purged from the solution. The LiOH trap         inorganic substances of the standard solution
            binds the CO 2 from the gas mixture                 are  converted  to  CO 2  and  transferred  to the
            (originating from the TIC). The volatile            NDIR detector using a carrier gas.
            compounds pass the trap and  reach the
            catalyst.