Page 10 - Clinical-Medical Solution for Clinical Chemistry
P. 10

Application   No. C153
    News


            „  Comparative Test with Manual Pretreatment       A comparison of quantitative values between the
                                                               methods  was performed for amiodarone and the
            A comparative test was performed between a manual
            pretreatment method and the fully automated        metabolite desethylamiodarone (Fig. 5 and 6, Table 4
                                                               and 5). In the wide range of concentrations detected
            pretreatment analysis method that employs the fully
            automated sample preparation LC/MS/MS system.      from the samples, there  was favorable agreement
                                                               between the quantitative results of the manual
            Human blood plasma for measuring the concentration
            of amiodarone was used.                            pretreatment method and the fully automated
                                                               pretreatment analysis method. The  coefficient of
            The manual pretreatment method involved manually   determination (R2) of both methods calculated from
            isolating the blood plasma, adding  acetonitrile, and   these results was 0.95 or higher (Fig. 7 and 8).
            mixing to perform deproteinization. After centrifugal   The fully automated pretreatment analysis method
            separation  of  this sample, the supernatant was then
            transferred to a vial for LC/MS/MS  analysis. On the   used by this system is a fast and low-burden analysis
                                                               technique that achieves quantitative results equivalent
            other hand, the fully automated pretreatment analysis
            method enabled the  entire analysis process, from   to conventional manual pretreatment methods and we
                                                               anticipate its utilization into the future.
            blood plasma isolation to  LC/MS/MS analysis, to be
            performed completely automatically using the system
            described in this article (Fig. 4).
































                       Pretreatment Workflow of the Manual Pretreatment Method and Fully Automated Pretreatment Analysis Method

                (x100,000)                                          (x100,000)
              1.0                                                1.0
                 1:AMD 646.00>58.10(+) CE: -52.0                    1:AMD 646.00>58.10(+) CE: -52.0
                 2:DEA 618.00>72.10(+) CE: -32.0                    2:DEA 618.00>72.10(+) CE: -32.0
              0.9                                                0.9
                                    Manual Pretreatment                              Automated Pretreatment
              0.8                                                0.8
                                                                                         with CLAM-2000
              0.7                                                0.7
                                        Amiodarone                                         Amiodarone
                                        546 ng/mL                                           557 ng/mL
              0.6                                                0.6
              0.5  Desethylamiodarone                            0.5  Desethylamiodarone
                       416 ng/mL                                          423 ng/mL
              0.4                                                0.4
              0.3                                                0.3

              0.2                                                0.2

              0.1                                                0.1
              0.0                                                0.0

             -0.1                                               -0.1
               1.50     1.75     2.00     2.25        min          1.50     1.75     2.00     2.25       min

                  Human Blood Plasma (Sample 3) Analysis Results Using   Human Blood Plasma (Sample 3) Analysis Results Using
                       the Manual Pretreatment Method               the Fully Automated Pretreatment Analysis Method
   5   6   7   8   9   10   11   12   13   14   15