Page 8 - Shimadzu SolidSpec-3700i
P. 8

Large Sample Compartment Accommodates a

            Wide Variety of Samples.




            The SolidSpec-3700i/3700i DUV have large sample     the sample compartment and an entire sample area of                   A wide variety of accessories, such as absolute     performed with the optional Automatic X–Y Stage
            compartments which allow large samples to be        12 inches or 310 × 310 mm is measurable by mounting                   specular reflectance attachments and relative        by inputting the intervals and the rotation angles
            measured without sample destruction. Their internal   the Automatic X–Y stage (optional). The vertical optical            specular reflectance attachments, expand the         for the sample.
            dimensions are 900W × 700D × 350H mm. A sample with   path makes it possible to perform transmission or                   application range. Automatic measurements can be
            a maximum size of 700W × 560D × 40H mm can be set in   reflectance measurements of large samples.
                                                                                                                                           Automatic Measurement
                 Large Sample Compartment
                                                                                                                                      Automatic X–Y Stage (Optional)
                                                                                                                                      The Automatic X–Y Stage developed for the SolidSpec-3700i/3700i DUV enables automatic measurements for the
                                                                                                                                      points specified in advance while maintaining the nitrogen gas purge.



















                                                                                                                                                12 inch Silicon Wafer on Automatic X–Y Stage           Input of Intervals and Rotation Angles


                                                                                                                                        Direct Measurement of Liquid Samples and Solid Samples without Integrating Sphere


                                                                                                                                      Direct Detection Unit (Optional)
                                                                                                                                      The SolidSpec-3700i/3700i DUV can measure samples   mounting the Direct Detection Unit (DDU-DUV) in the
                                                      Large Sample Compartment                                                        with an integrating sphere as the standard detection   SolidSpec-3700i DUV, measurements down to 165 nm (note)
                                           A sample of 700W × 560D mm is set in the sample compartment.                               system. However, some samples require measurement   are possible. Measurements with the Direct Detection
                                                                                                                                      without using an integrating sphere. The Direct     Unit can be performed simply by switching a mirror.
                                                                                                                                      Detection Unit was created for just such instances. By
                  Three Dimensional Optical Path

            The three-dimensional optical path enables
            non-destructive measurement of large samples, without
            having to cut them smaller. In the optical path of previous
            models, light only traveled horizontally, but the new
            models include a three-dimensional optical path (U.S.
            patent 6583872) with light also traveling in the vertical
            direction. Samples can be placed horizontally, which
            makes it easier to place large samples.

                                                                                                                                              Measurement of Film Sample with Direct Detection Unit  Measurement of Liquid Sample with Direct Detection Unit


                                                                                                                                      Note) In order to measure below 190 nm with SolidSpec-3700i DUV, nitrogen gas purging is required to remove interfering oxygen molecules inside of
                                                                              Three Dimensional Optical Path                              the SolidSpec-3700i DUV. The measurable range for the SolidSpec-3700i with the optional Direct Detection Unit is 190 nm to 3300 nm.

                                                                                                                                                                                                      SolidSpec-3700i/3700i DUV
      8                                                                                                                                                                                                                     UV-VIS-NIR Spectrophotometer  9