High Sensitivity



 High accuracy for transmittance and reflectance is required for the measurement of optical parts.   Comparison of Two-Detector and Three-Detector Models
 The SolidSpec-3700i/3700i DUV have three detectors which cover the range from ultraviolet to near-infrared.
 The sensitivity in the near-infrared region is significantly enhanced by using both InGaAs and cooled PbS   An InGaAs detector is used, in addition to the photomultiplier tube (PMT) and cooled PbS detectors.
 detectors. Highly accurate and highly sensitive spectra are obtainable from ultraviolet to near-infrared.  That results in less noise than the two-detector (PMT and PbS) model, especially in the region detected by the InGaAs
             detector (900 to 1600 nm).

 Three Detectors     1.000                                    70.000

                                                              60.000




                                                              40.000
                     0.500
                    T%                                       T%
 PbS detector

                                                              20.000

 InGaAs detector
                     0.000
                                                               0.00
                    −0.100                                    −5.000
                      1000.00  1200.00  1400.00  1600.00       1000.00  1200.00  1400.00  1600.00  1800.00  2000.00
 Three detectors attached to the integrating sphere  nm                       nm
                     : SolidSpec-3700i/SolidSpec-3700i DUV (InGaAS detector)  The spectrum above was measured with InGaAs and PbS
 PMT detector        : 2-detector model (PbS detector)       detectors, using a sharp cut filter for near-infrared light. Increasing
                                                             the sensitivity in the near-infrared region is especially beneficial for
                                                             low-reflectance measurements, such as measuring anti-reflective
 Sensitivity Characteristic                                  (AR) coatings used in the optical communications field.
 100  Conventional spectrophotometers have used a PMT (photomultiplier
 Relative Value (%)  PMT  tube) for the ultraviolet and visible region and a PbS detector for the   High Accuracy Measurement with Minimized Detector Switchover

            Noise and Bump
 InGaAs
 near-infrared region. However, neither detector has much sensitivity
 PbS
 near the wavelength of 900 nm, preventing high-sensitivity
 measurement in this range.   Noise and bump caused by detector switchover are minimized to assure accurate measurement.
 The SolidSpec-3700i/3700i DUV make it possible to take
 0
 1000  2000  3000  high-sensitivity measurements in the switchover range by using an   30.000  100.000
 Wavelength (nm)
 InGaAs detector as shown in the figure on the left.
                   20.000
                                                             90.000
 Relationship between Detectors and Measurable Range
                   T%                                        T%
 165 nm  380 nm  780 nm  3300 nm
                   10.000
 UV  Visible  NIR                                            80.000
 PMT  165 to 1000 nm
 InGaAs  700 to 1800 nm  0.000
 PbS  1600 to 3300 nm  -3.000                                70.000
                     700.00   800.00   900.00   1000.00       1550.00  1600.00  1650.00  1700.00  1750.00
                                    nm                                        nm
 The photomultiplier tube detector can be switched to the InGaAs detector in the range from 700 nm to 1000 nm (the
 default switching wavelength is 870 nm). The InGaAs detector can be switched to a PbS detector in the range from 1600 nm   The transmission spectra of a color filter and polyester film are shown in the left and right figures, respectively.
 to 1800 nm (the default switching wavelength is 1650 nm).  Noise or bump caused by the detector switchover range at 870 nm and 1650 nm are not observed.


                                                                            SolidSpec-3700i/3700i DUV