Wide Measurement Wavelength Range (SolidSpec-3700i DUV)



             The development of precise laser machining using an ultraviolet laser such as an ArF excimer laser enhances              Integrating Sphere for Deep Ultraviolet Measurement
             the requirement for transmittance or reflectance measurements of optical parts in the deep ultraviolet region.
             The SolidSpec-3700i DUV (note 1)  enables measurement in the range of 175 nm to 2600 nm (note 2)  with an integrating
             sphere and the range of 165 nm to 3300 nm (note 3)  by mounting the optional Direct Detection Unit DUV. With this           150.000
             additional unit, the range from deep ultraviolet to near-infrared is now measurable.                                                                                         Spectra with low noise can be obtained even for wavelengths
                                                                                                                                         140.000
                                                                                                                                                                                          near 190 nm, which are difficult to measure accurately using a
                                                                                                                                                                                          regular spectrophotometer.
                                                                                                                                                                                          The ability to measure spectra in this wavelength range is
                 Nitrogen Gas Purge                                                                                                      120.000                                          especially helpful when measuring semiconductor materials used
                                                                                                                                                                                          for ArF excimer lasers.
                                                                                Oxygen molecules in the atmosphere absorb               T%
                                                                                ultraviolet light under 190 nm. Nitrogen gas             100.000
                                                                                purging for both the optical and the sample
                                                                                compartment is required to remove the
                                                                                interfering oxygen molecules. Since the                  80.000
                                                                                SolidSpec-3700i DUV has purge inlets for
                                                                                each compartment, efficient nitrogen gas
                                                                                purge is possible so that the time required for          60.000                                     The 100% baseline spectra measured on the SolidSpec-3700i DUV with an
                                                                                purging after sample replacement is reduced,                                                        integrating sphere for deep ultraviolet measurement and the SolidSpec-3700i
                                                                                                                                                                                    with a normal integrating sphere are shown in the left figure.
                                                                                and high sensitivity with lower stray light in           50.000   200.00       250.00       300.00
                                                                                                                                           175.00
                                                                                the deep UV region is achieved.                                              nm
                                                                                                                                               : SolidSpec-3700i DUV with an integrating sphere for the deep
                                                                                                                                                 ultraviolet measurement
                                                                                                                                               : SolidSpec-3700i with a normal integrating sphere





                                                                                                                                      Example of Deep Ultraviolet Region measurement


                                                                                                                                         70.000
                Integrating Sphere and Photomultiplier for the Deep Ultraviolet Region                                                                                                    In order to perform high-accuracy measurements in the deep UV
                                                                                                                                                                                          region, a sufficient quantity of light and significant low stray
                                                                                                                                         60.000                                           light are required. The transmission spectrum of a silica plate
                                                                                Materials which do not absorb deep                                                                        measured with the Direct Detection Unit DDU-DUV (optional) is
                                                                                ultraviolet light are required to be used as                                                              shown in the left figure. Spectra with significantly lower noise
                                                                                the window material for the detector and                                                                  are obtainable in the ultraviolet region.
                                                                                the material for the inside of the
                                                                                                                                         40.000
                                                                                integrating sphere to enable performance                T%
                                                                                in the deep ultraviolet region. The
                                                                                SolidSpec-3700i DUV or D 2 lamp uses a
                                                                                PMT detector with fused silica as the
                                                                                window material and an integrating                       20.000
                                                                                sphere with resin that has highly reflective
                                                                                characteristics in the deep ultraviolet
                Integrating Sphere for Deep Ultraviolet  Photomultiplier for Deep Ultraviolet  region as the inside material.
                                                                                                                                          0.000
                                                                                                                                           165.00 170.00  180.00   190.00   200.00  Transmission Spectra of Silica Plate
                                                                                                                                                             nm
            Note 1) In order to measure the range below 190 nm with the SolidSpec-3700i DUV, nitrogen gas purge is required to remove interference from oxygen molecules inside the SolidSpec-3700i DUV.   : Transmission spectrum of Silica plate measured with nitrogen purge
            Note 2) The measurable range for SolidSpec-3700i is 240 nm to 2600 nm.
            Note 3) The measurable range for SolidSpec-3700i with the optional Direct Detection Unit is 190 nm to 3300 nm.                : Transmission spectrum of Silica plate measured without nitrogen purge