Page 9 - Shimadzu MXF-2400
P. 9

7  Enhanced Sensitivity for Superlight  9  Analytical Information can be set for
 Elements
                  Each Individual Sample Form and Each
                  Individual Element
 The layered structure analyzer (LSA), which uses curved crystals
 for convergence, provides almost ten times higher sensitivity for   Analytical conditions such as optimal X-ray output and PHA
 X-rays than conventional detectors.  LSA dispersing element  Monochromator for LSA  (pulse height analyzer) range can be freely set for each
 Unlike the conventional dispersing crystals, the LSA has   individual sample form and element.
 synthetic multilayer structures.  Also, an automatic sensitivity correction function (counting loss
 The spacer is made of the material and has the thickness that   12  TAP. (Mg, Na, F)  correction) can be used.
 gives the best lattice constant to sense X-ray intensities at the   11  LSA dispersing element
 highest sensitivity for such elements as Be, B, C, N, O, F, Na, and   10
 Mg.  9
 X-ray intensity ratio  8 7 6






 4 5
 3
 2
 1
 Mg  Na  F  O  N  C  B  Be
 Element
 X-ray Intensity Comparison between LSA Dispersing Element
 and TAP Crystal Total Reflection




 8  Wide Dynamic Range  ( x 10  CPS)
 4
 X-ray tube voltage: 40 kV
 The Shimadzu’s original pulse counting circuit and the counting
 300  X-ray:  CuKa
 error correcting circuit combine to provide a wide dynamic   Simultaneous Determination of up to 36
 range: one calibration curve can cover the concentration range   10 Major and Impurity Elements
 more than 3,000,000 cps, from 0% to 100%. High accuracy is
            The 36 fixed monochromators detect impurity elements as well
 ensured even in high concentration region where the X-ray
 X-ray intensity  few ppm concentration level.
 intensity is very high.  200  as major elements with high reliability. The detection limit is a
            The scanning monochromator permits automatic qualitative
            determination and may also be used for quantitative
            determination.
 100
            There is a strong demand for the method to analyze not only
            major elements but also impurity elements, in order to enhance
            the quality of the final products. The technique of X-ray
            fluorescence spectrometry has detection limits of
 0  10  20  30  40  50  60  a few ppm concentration level and is applicable to various types
 Linear Calibration Curve more than 3,000,000 cps  X-ray tube current (mA)  of samples.







                                                                                                   MXF-2400
                                                                                           Multi-Channel X-ray Fluorescence Spectrometer
   4   5   6   7   8   9   10   11   12   13   14