Page 5 - Shimadzu MXF-2400
P. 5

Principle   Construction








 When X-rays hit a sample, atoms in the sample are excited and release the secondary X-rays, which are also called fluorescent X-rays.
 Since the wavelength of secondary X-rays is peculiar to the element concerned, the sample can be qualitatively determined by measuring
 that wavelength. Also, since the fluorescent X-ray intensity is proportional to the concentration of the element, quantitative
 determination is made by measuring the fluorescent X-ray intensity of the wavelength peculiar to each element.
                                                               1

 Focusing circle
 Detector  X-ray tube
 Detector
 Secondary slit                                                    2
 Secondary slit                                           3
 2θ 2
 2θ 2                                      6
 Curved crystal
 Curved crystal                                       4                       7
 Primary slit
 Primary slit
 Sample
 Concentration
 Concentration (%)  calculation   X-ray intensity (kcps)                       8
 Si  3.1182  Si  34.79725
 Mn  0.2011  Intensity  Mn  6.30588
 P  0.0209  P  0.45881                                       5
 S  0.0173  S  1.11912
 Mg  0.0185  Concentration  Mg  0.11576
                                                                               9


 Spectrometer Unit  Measuring Electronics

 This unit disperses the fluorescent X-rays generated from the   The electric pulses corresponding to the number of X-ray photons
 sample, measures the intensity of the X-rays of the particular   are counted and recorded. — The output pulse signal of the
             1  X-ray Spectrometer (Installed in the case)      6  Data Processing Unit
 wavelength, and converts that intensity into electric signals. High   detector is amplified and its interfering spectrum signals are
 quality curved crystals are used for dispersion.   removed by the pulse height analyzer. Then the signal is measured   Up to 36 monochromators are radially arranged around   The operation of the MXF-2400 is made via the keyboard
            the X-ray tube. The case of the X-ray spectrometer is   of this unit. This unit has various high-level data handling
 The MXF-2400 uses one monochromator for one element, and up   by the scaler, processed by the microprocessor, and then
            temperature-controlled.                            functions, which provides advanced analysis by easy
 to 36 monochromators can be installed together to permit   transmitted to the data processing unit.
                                                               operation.
 simultaneous determination of up to 36 elements.
             2  Sample Turntable (with dust-proof cover)
 Use of the optional scanning monochromator provides automatic   7  X-ray Power Control Unit and
 qualitative analysis and allows up to 30 elements to be preset for   Up to eight samples may be loaded together for   Processor Unit
 quantitative analysis.  automated successive analysis. An external automatic
            sample feeder and/or automatic sample pretreatment unit   The X-ray power control unit controls the output power
            may be connected to this turntable.                of the X-ray tube with such a high stability that it is not
 X-ray Power Source  Data Processing Unit                      necessary to use a motor generator in most cases. The
             3  Sample Feeding Unit                            processor unit incorporates the X-ray intensity measuring
 This unit supplies power to the X-ray tube which emits the   X-ray intensities are converted into concentration values of the
               (Installed behind the panel)                    circuit and the microprocessor, which counts X-ray pulses
 excitation X-rays (primary X-rays). It consists of a high-voltage   elements concerned through the use of the conversion equations   and controls all the other units.
 transformer and an X-ray power controller.  (calibration curves) predetermined using standard samples. The   The swing arm system accurately positions the sample.
 results are presented on the display or the printer.           8  DC Power Source Unit
             4  Maintenance Panel
                                                               This unit supplies DC power to the control units.
 Spectrometer   This unit displays the status of each part and has manual
 unit
            switches for checking.
 Measuring electronics  Data processing unit
 Mono-                                                          9  Switchboard
 Microprocessor  Keyboard  5  X-ray Tube Cooling Unit          The necessary power source is one three phase 200 VAC
 chro-  Display
 X-ray power source  meter  Detector
 Printer
 PHA scaler  This unit circulates cool, distilled water to the X-ray tube.   or 220 VAC line.
 X-ray tube  Amplifier
            The distilled water is cooled externally.
                                                                                                  MXF-2400
                                                                                           Multi-Channel X-ray Fluorescence Spectrometer
   1   2   3   4   5   6   7   8   9   10