Page 7 - Shimadzu MXF-2400
P. 7

Features








 1  The World's First Multi-Channel  P  S  4  The Effect of Nearby or Interfering  100  Scintillation counter  Kr exatron  Gas flow proportional counter
 Fluorescent X-ray Spectrometer to be
                  Elements can be Reduced
                                                                                                 (5µ polypropylene)
 Installed with a 4 kW (Thin Window)  1.5  Mn  Ni  : X-ray intensity  Mo  80  Kr multitron  Ar exatron  Gas flow proportional
                                                                                             counter (6µ Mylar)
 X-ray Tube  A detector suitable for each element can be selected from between   Counting efficiency (%)  60  Xe exatron
            the gas sealed detector sealed with gas appropriate for elements such   Ar multitron       Ne exatron (Al)
 Sensitivity has been enhanced to approximately 1.3 times for heavy   as Ne, Ar and Kr (selection can be made over a broad range from   40  Ne exatron (Be)
 elements and 1.7 times or more for light elements (elements lighter   1.0  : BEC  11Na to 92U),  Ne multitron
                                                                 20
 than Cl).  the scintillation detector, and the gas flow detector.
 : Detection limit  The effect of nearby or interfering elements can be minimized as the   0  1  2  3  4  5  6  7  8  9  10  11  12
 0.7                                                            K line
 15  25  35  45  optimum detector can be selected to provide excellent detection   BaMo  Cu Fe Cr Ti  Ca K  Cl  S  P  Si  Al  Mg  Na
 Atomic Number  efficiency and good resolution for each element.  L line  UPbW  CeBa  SnCd  Mo
                                                                               Counting efficiency of detectors

 2  Excellent Analytical Accuracy from  2θ  Secondary slit  Excellent Measuring Reproducibility for
 Minute Regional Areas to Whole Areas

 The converging spectrometer system is one of Shimadzu’s original   5  Light Elements  100
 designs. The fluorescent X-ray focus doesn't be diverged, and
            Using Shimadzu’s original vacuum stabilizer (patented), a constant
 becomes extremely small at the secondary slit so that reflected X-ray
 Curved crystal  vacuum can be maintained during measuring. And as the change in
 intensity is extremely high, and resolution is good in comparison with
            X-ray intensity due to the change in the degree of vacuum (change   Degree of vacuum Pa
 other systems.
            due to air absorption) can be suppressed, excellent measuring
            reproducibility can be achieved for light elements (such as Al  and Si)        Vacuum stabilizing point
            and superlight elements (such as Be, B and C).
                                                               Approx. 15                         With vacuum stabilizer
 Primary slit                                                                                     Without vacuum stabilizer

 3  Excellent Long-term Stability and                                            The point of starting analysis  Time (sec.)
 Maintenance-free Characteristics

 Through the development of original technology by Shimadzu an
 excellent gas sealed detector (SPC) is used to obtain long-term stability   Exatron for detection of
 heavy and light elements
 and maintenance-free characteristics over a wide range (11Na to 92U) of   Excellent Long-term Stability for
 light elements including Na which normally only could be analyzed   6  Superlight Elements (Be, B, C, N, O, F)
 using a gas flow detector (FPC).
 Shimadzu’s own original gas sealed detector has the following   The gas flow detector (FPC) has excellent long-term stability as pulse
 features.  height distribution is stabilized through the use of an automatic core
 1) Starting up X-ray signals is fast, and a large number of X-ray signals   wire winding system (Shimadzu patent) and a gas density stabilizer
 can be counted at a high speed.  with CPU control system.
 2) CO2 is mixed in with the inert gases (Ne, Ar, Kr) to give good gas   The automatic core wire winding system is different to the previous
 Multitron for detection of   Scintillation counter for
 stability, and the core wire does not become contaminated as there   heavy elements  heavy elements  core wire cleaning system and manual core wire winding system in
 is no resolution deterioration due to ionization.  that core wire winding is continuously automatic (5 mm per day) so
 3) As there is no gas piping, consideration of the surroundings is not   %  CaO  Statistical error 1σ (C.V. 0.084%)  that the core wire does not become contaminated, analysis can be
 Concentration
 +0.4  44.7%                                                                Anode           Window for X-rays
 required, and excellent long-term stability is assured. (The device is   +0.2  0.084%  conducted always with a new core wire, and maintenance is not
 0
 ideal for process control and automatic analysis as the number of   –0.2  needed.
 –0.4
 calibration curve adjustments and α/βcorrection can be minimized.)  Furthermore, the PR gas flow and density necessary for the gas flow
 4) As the gas is sealed in the device, there is no maintenance, which   %  Fe2O3  Statistical error 1σ (C.V. 0.22%)  detector achieves stability in the gas density stabilizer with CPU   PR gas
 Concentration
 makes the product extremely permanent. Moreover, the whole   +0.8  2.1%  0.22%  control system.
 +0.4
 spectrometer unit is kept stable as a thermostat in the device   0
 –0.4
 continually maintains the temperature to within ±0.2°C of the set   –0.8
                                                                               New core wire  Used core wire
 temperature.  (C.V)  1st day  5th day  10th day  15th day  20th day  25th day  30th day
                                                                                                  MXF-2400
                                                                                           Multi-Channel X-ray Fluorescence Spectrometer