Page 14 - Shimadzu LCMS-8060
P. 14

Engineered for Robustness and

            Easy Operation/Maintenance
                                                                                                                                      Newly Designed Ionization Unit


            The robustness of the LCMS-8060 and modified ion optics was   a deuterated internal standard (alprazolam-d5) the RSD was 3.5%.   Designed without cables or tubes, removing the new ionization unit is simple: release a one-touch lever to open the unit and lift it out. In
            also assessed by injecting 2400 samples of femto-gram levels of   As part of the robustness test the vacuum system was vented to   addition, no tools are needed to detach the needles fitted in APCI and DUIS  units, allowing for easy maintenance.
                                                                                                                                                                                             ™
            alprazolam spiked into protein-precipitated human plasma extracts   model a transient power failure with no effect on signal response
            over a 6 day period (over 400 samples were injected each day).   or baseline noise level.
            The RSD of peak area response was 5% over this test period, using

                          ×10 4
                        7.0                                                                       1.4
                           100 pg/mL alprazolam in human plasma with internal standard
                           m/z 309.05 > 281.10
                        6.0                                                                       1.2
                              1st day     2nd day    3rd day       4th day     5th day    6th day
                        5.0                                   A week                              1.0
                       Peak Area  4.0                  Area Ratio 3.46 %RSD                       0.8  Area Ratio
                                                       Area Ratio 3.46 %RSD
                                                                                                  0.6
                        3.0
                        2.0                                                                       0.4
                                                       Peak Area 4.97 %RSD
                        1.0                                                                       0.2
                                                            Vacuum
                                   Pause in batch  Pause in batch  system  Pause in batch  Pause in batch
                                    analysis    analysis    vented       analysis    analysis
                        0.0                                                                       0.0
                          0          400         800         1200        1600        2000       2400
                                                          Injection number
                                                 Long term stability study on LCMS-8060
                           (×10,000)                 (×10,000)                (×10,000)
                         1.00                      1.00                     1.00
                            1 st  injection          1200 th  injection        2400 th  injection
                         0.75                      0.75                     0.75
                         0.50                      0.50                     0.50
                         0.25                      0.25                     0.25
                         0.00                      0.00                     0.00
                          0.8  0.9  1.0  1.1  1.2  1.3 min  0.8  0.9  1.0  1.1  1.2  1.3 min  0.8  0.9  1.0  1.1  1.2  1.3 min
                                  MRM chromatograms for the 1st, 1200th and 2400th measurements of alprazolam
                                                                                                                                                   Steps for ESI Capillary Replacement
                                              Intraday and Interday variations on LCMS-8060
                                                     Intraday Variation (%RSD)       Interday Variation (%RSD)
                            Compound
                                         1st day  2nd day  3rd day  4th day  5th day  6th day  Days 1–3  Days 4–6  6 Day Total                                                      3
                                                                                                                                                        1
                            Alprazolam   5.04   4.94   5.06   5.38   4.55   4.83   3.19   1.63   2.74
                                                                                                                                                                                   2
                         Alprazolam-d5 (ISTD)  5.04  4.68  5.48  5.31  4.26  4.91  2.62   1.89   2.18
                                                                                                                                                                                                                       ESI Capillary
                            Area ratio
                       (Alprazolam / Alprazolam-d5)  3.48  3.11  3.48  3.44  3.71  3.54  1.79  0.26  1.40

            Easy System Maintenance Reduces Downtime                                                                                   ESI (standard)                  APCI (optional)                   DUIS (optional)

            As with Shimadzu’s other triple quad systems, maintaining the LCMS-8060 is simple. Replacing the desolvation line (DL) and ESI capillary
            is quick and easy. Additionally, the design allows users to replace the DL without breaking vacuum, providing greater uptime and usability.

                            Steps for DL Replacement

                                                           2
                           1

                                                               3
                                                                                      DL (Desolvation Line)



      14                                                                                                                                                                                                   Liquid Chromatograph Mass Spectrometer  LCMS-8060  15