Page 8 - Shimadzu ICPMS-2030 Series
P. 8

Higher Sensitivity and Lower Interference






                     Achieve High-Sensitivity and Low-interference

                     Analyses Using Newly Developed Collision Cell



                     With this newly developed collision cell,                         Newly Developed Collision Cell
                     polyatomic ions that previously may have
                     been detected as spectral interference are
                     ef ciently eliminated by helium gas  owing
                     through the cell, allowing the
                     high-sensitivity analysis.







                     Quantitative Analysis of  As           Without Collision Cell        With Collision Cell
                                            75
                     If chlorine is in the sample when performing
                     quantitative analysis of  As,  Ar and  Cl   75 As                                75 As
                                       40
                                             35
                                    75
                                                            in HCl 1%                                 in HCl 1%
                                   35
                                40
                     combine to form  Ar Cl, which overlaps the   Blank                               Blank
                     mass number of As, creating interference.  in HCl 1%                             in HCl 1%
                                                                           35
                                                                         40 Ar Cl interference eliminated
                                                                         Highly sensitive analysis of  As
                                                                                        75
                      The collision cell eliminates the spectral
                      interference from Cl, enabling the highly sensitive
                      measurement of As.                      40 Ar 35 Cl
                     On-Line IEC (Inter-Element Correction)



                     Shimadzu’s proprietary On-Line IEC (Inter Element Correction) provides correction for spectral interference that could not be eliminated
                     with the collision cell. Since the interference correction coef cient can be acquired only by a single measurement of a standard sample
                     of the interfering element, the correction can be simply and accurately made. There's no need for complicated assessments of
                     conditions, as with reaction cell technology.


                                            78
                     Quantitative Analysis of  Se             By measuring a standard sample of Gd, the correction coef cient can be obtained.
                                                              Then, the portion where interference occurs can be corrected when measuring
                     If  Gd is in the sample when performing quantitative analysis   the unknown sample, and the accurate quantitation of  Se can be accomplished.
                      156
                                                                                             78
                     of  Se, spectral interference will occur with respect to  Se.
                                                       78
                       78
                         Measurement of      Calculation of      Analysis of  Calculation of amount of
                       Gd standard sample  correction coef cient  unknown sample  156 Gd interference on  Se  Correction
                                                                                            78
                            A
                        α =
                            B           Correction coef cient: α                      A′ = B′ × α
                                     B       Ratio of  Gd to
                                                  156
                        A                  interfering component                A′          B′
                                              is constant.
                           78 Se  156 Gd                        78 Se  156 Gd      78 Se  156 Gd       78 Se
                      Interference of doubly charged            78 Se containing   Components       Accurate  Se
                                                                                                         78
                                                                                          78
                            156
                                      78
                       ion from  Gd appear to  Se                 interference  interfering with  Se
        8
   3   4   5   6   7   8   9   10   11   12   13