Standard Height W 798 × D 515 × H 1582 mm W 975 × D 579 × H 1708 mm W 1206 × D 765 × H 2170 mm
+250 mm Extension W 798 × D 515 × H 1832 mm W 975 × D 579 × H 1958 mm W 1206 × D 765 × H 2420 mm
Dimensions* 6
+500 mm Extension W 798 × D 515 × H 2082 mm W 975 × D 579 × H 2208 mm W 1206 × D 765 × H 2670 mm
+750 mm Extension — — W 1206 × D 765 × H 2920 mm
100 V Model 210 kg — —
Standard 200 V Model 210 kg 410 kg 660 kg
Height
400 V Model — — 670 kg
100 V Model 230 kg — —
+250 200 V Model 220 kg 420 kg 670 kg
Extension
Weight 400 V Model — — 680 kg
100 V Model 240 kg — —
+500 200 V Model 230 kg 440 kg 690 kg
AGX-10kNV2D AGX-50kNV2D AGX-50kNV2 Extension 400 V Model — — 700 kg
100 V Model — — —
System Specifications +750 200 V Model — — 700 kg
Extension
400 V Model — — 710 kg
Model Name AGX-10kNV2D AGX-20/50kNV2D AGX-20/50kNV2 100 V Model 1.5 kVA/15 A — —
Max. Loading Capacity 10 kN 50 kN (Single-Phase, 100 to 115 V)
Loading Method AC servomotor drive 200 V Model
Power Supply/ 2.0 kVA/10 A 5.5 kVA/30 A —
Within ±0.5 % error of the indicated test force (in a range from 1/1000 to 1/100 of the load cell rated capacity) (Single-Phase, 200 to 230 V)
High-Accuracy Type Within ±0.3 % error of the indicated test force (in a range from 1/100 to 1/1 of the load cell rated capacity) Breaker Capacity 200 V Model
Compatible with JIS B7721 Class 0.5, EN 10002-2 Grade 0.5, ISO 7500-1 Class 0.5, BS 1610 Class 0.5, DIN 51221 Class 0.5, and ASTM E4 Required — — 4.5 kVA/15 A
(3-Phase, 200 to 230 V)
Test Force Within ±1 % error of the indicated test force (in a range from 1/2000 to 1/1000 of the load cell rated capacity) 400 V Model
Measuring Within ±0.5 % error of the indicated test force (in a range from 1/1000 to 1/100 of the load cell rated capacity) (3-Phase, 380 to 400 V) — — 4.0 kVA/10 A
Range* 1 Wide Ranging Type Within ±0.3 % error of the indicated test force (in a range from 1/100 to 1/1 of the load cell rated capacity)
Compatible with JIS B7721 Class 1, EN 10002-2 Grade 1, ISO 7500-1 Class 1, BS 1610 Class 1, DIN 51221 Class 1, and ASTM E4 *1: In the JIS B7721, EN 10002-2, ISO 7500-1, and ASTM E4 standards, an inspection is recommended after the testing machine is installed.
*2: The crosshead speed accuracy is calculated from the amount of crosshead movement within a specified time at a crosshead speed of 0.5 to 500 mm/min under normal conditions.
Within ±1 % error of the indicated test force (in a range from 1/1000 to 1/1 of the load cell rated capacity)
Compatible with JIS B7721 Class 1, EN 10002-2 Grade 1, ISO 7500-1 Class 1, BS 1610 Class 1, DIN 51221 Class 1, and ASTM E4 *3: The tensile stroke indicates the value when manual non-shift wedge grips (MWG) are attached.
Standard-Accuracy Type *4: Use the voice operation function when the noise is 55 dB(A) or less. Voice operations may not work when the background noise is high.
Within ±1 % error of the indicated test force (in a range from 1/500 to 1/1 of the load cell rated capacity) *5: 10 N and 20 N are only for the standard-accuracy type.
Compatible with JIS B7721 Class 1, EN 10002-2 Grade 1, ISO 7500-1 Class 1, BS 1610 Class 1, DIN 51221 Class 1, and ASTM E4 *6: During installation, ensure a space of 600 mm on each side and at the rear of the testing machine for maintenance.
• Values in this catalog have been measured based on separately prescribed test standards.
Crosshead Speed Range* 2 0.0005~3000 mm/min 0.0005~1500 mm/min 0.00005~1500 mm/min • If small conductive sample fragments are produced, they might get inside the main unit, resulting in malfunctions. In such a case, a separately installed controller is recommended.
Maximum Crosshead Return Speed 3000 mm/min 2000 mm/min
Crosshead Speed Accuracy ±0.1 %
Crosshead Speed and Permitted Test Force Up to the maximum load capacity for all speed ranges
Crosshead Measurement System Battery-less multi-turn absolute encoder
Position Detection Positional Accuracy Within ±0.05 % of the indicated value, but ±0.01 mm when the indicated value is below 20 mm Options
Crosshead Position Control Resolution 12.5 nm 8.33 nm
P/N Description
180 to 1150 200 to 1150 215 to 1265
Standard Height 339-90000-01 Exclusive Table, Height 650 mm, for the 10 kN Max. Table-Top Type
(0 to 550) (20 kN:0 to 750 mm, 50 kN:0 to 690 mm) (20 kN:0 to 860 mm, 50 kN:0 to 800 mm)
339-90000-02 Exclusive Table, Height 400 mm, for the 10 kN Max. Table-Top Type
180 to 1375 200 to 1375 215 to 1490
Crosshead-Table +250 mm Extension (0 to 775 mm) (20 kN:0 to 975 mm, 50 kN:0 to 915 mm) (20 kN:0 to 1085 mm, 50 kN:0 to 1025 mm) 339-90001-01 Exclusive Table for the 20/50 kN Table-Top Type
Clearance (mm) 336-00313-01 Anti-Topple Attachment for the 10 kN Max. Table-Top Type
(Tensile Stroke)* 3 +500 mm Extension 180 to 1600 200 to 1600 215 to 1715 336-00313-02 Anti-Topple Attachment for the 20/50 kN Max. Table-Top Type
(0 to 1000 mm) (20 kN:0 to 1200 mm, 50 kN:0 to 1140 mm) (20 kN:0 to 1310 mm, 50 kN:0 to 1250 mm)
336-00311 Anti-Topple Attachment for the 20 to 300 kN Floor Type
215 to 1940
+750 mm Extension — — 336-00311-01 Anti-Topple Attachment for the 300 kN +750 Extension Floor Type
(20 kN:0 to 1535 mm, 50 kN:0 to 1475 mm)
336-00311-02 Anti-Topple Attachment for the 600 kN Floor Type
Effective Test Width 420 mm 500 mm 600 mm
336-01076-01 Sensor Amplifier
Data Capture Rate 10 kHz max. 336-01076-07 Analog Input Amplifier
Direction of Tension 60 kN/mm 180 kN/mm
Frame Rigidity 336-01076-04 Analog Output Amplifier
Direction of Compression 60 kN/mm 180 kN/mm 336-01076-05 Counter Unit
• Test force/stress value display • Touch load detection • Test piece protection 336-01076-03 PIO Unit
• Stroke display • Switching display languages to • Timer 336-01076-02 Insulated PIO Unit
• Test force auto zero Japanese/English/Chinese language • Calculator 336-01076-08 Analog Recorder Unit
• Test force auto calibration • Switching display unit • Light 336-01076-06 Strain amplifier
• Automatic loading of load cell characteristic values • Standby Power Savings • Sound (Choose between standard and clear) 345-05842 X-TP Recorder
• Fine adjustment of crosshead position (Button/Dial) • Self-check • Audio Output (Japanese/English/Chinese) 345-05843 X-TYP Recorder
Standard Functions • Interlock function (Safety cover) • Auto return • 6 internal unit ports
When the displacement gauge is connected • Setting Jig distance (One of the ports is dedicated to testing power.) 346-55042 One-Touch Load Cell Replacement Function
• Displacement/strain value display function • Voice operation function *4 346-55042-01 One-Touch Load Cell Attachment
• Displacement auto zero When the operation controller is connected 336-01674-01 Safety Cover Offset 10kND STD
• Displacement auto-calibration • Single test control function • Cycle test control function 336-01674-02 Safety Cover Offset 10kND +250
(Displacement gauge input amplifier only) • Stress automatic test control function 336-01674-03 Safety Cover Offset 10kND +500
• Automatic test force/strain control (with autotuning function) 336-01674-11 Safety Cover Offset 50kND STD
• Breakage detection • Soft limit detection 336-01674-12 Safety Cover Offset 50kND +250
• Peak value/breakage value display • Crosshead speed preset function 336-01674-13 Safety Cover Offset 50kND +500
• Test condition internal memory file function (25 files) • S-S curve display function 336-01674-21 Safety Cover Offset 50/100kN STD
• USB flash drive connection function (window capture/real-time data sampling)
336-01674-22 Safety Cover Offset 50/100kN +250 / 300kN STD
• Up to 5 optional units listed below can be installed. 336-01674-23 Safety Cover Offset 50/100kN +500 / 300kN +250
Sensor amplifier (load cell, SG displacement meter, LVDT displacement meter), analog input amplifier (4CH), analog output amplifier (4CH), 336-01674-24 Safety Cover Offset 50/100kN +750 / 300kN +500
Optional Functions strain amplifier (2CH), counter unit (4CH), PIO unit (16 inputs and 16 outputs), insulated PIO unit (16 inputs and 16 outputs), analog recorder unit 336-01674-25 Safety Cover Offset 300kN +750
• Pneumatic or hydraulic grips interlocking operation 336-01073-13 USB camera stand 50 to 300 kN camera included
Standard Accessories Load cell, CAL cable, Tool set, Power supply cable, Rotating rod, Hexagon wrench, Instruction manual, Safety caution sheet (1 each) 336-01073-12 USB camera stand 50 kND camera included
Model Lineup by Load Cell Capacity* 5 10 N/20 N/50 N/100 N/500 N/1 kN/5 kN/10 kN 20 kN/50 kN 336-01073-11 USB camera stand 10 kND camera included