Page 101 - Shimadu Consumables Catalog
P. 101
Shim-pack ISA/ISC Series
Shim-pack ISA/ISC series uses polystyrene gel as solid support, making it possible to utilize both electrostatic reaction and
hydrophobic reaction. They are suitable for the analysis of sugars (ISA) and guanidiono compounds (ISC-05).
Peaks
1. Sucrose 7. Mannose
2. Cellobiose 8. Fructose
3. Maltose 9. Galactose
4. Lactose 10. Xylose
5. Rhamnose 11. Giucose
6. Ribose
Conditions
Column : Shim-pack ISA-07/S2504 (250 mmL × 4.0 mmI.D., 7 μm) (P/N: 228-09699-91)
Mobile Phase : Potassium borate buffer solution
gradient elution
Flow Rate : 0.6 mL/min
Col. Temp. : 65 °C
Detection : RF EX. 348 nm, Em. 430 nm
(Post-column derivatization with arginine)
Analysis of Saccharides Standards
Product Information
Particle Size Dimensions
Column Stationary Phase P/N Guard Column
(μm) (Length × I.D., mm)
Shim-pack ISA-07/S 2504 Quaternery ammonium group 7 250 × 4.0 228-09699-91 228-00823-91
Shim-pack ISC-05/S 0504 Na type sulfone group 5 38 × 4.6 228-09700-91
228-00802-91
Shim-pack ISC-07/S 1504 Na type sulfone group 7 150 × 4.0 228-09328-91
Shim-pack ISC-07/S 1504 Li Li type sulfone group 7 150 × 4.0 228-00796-91 228-00797-91
Check out more Application Systems at
www.shimadzu.com/an/hplc
Pg A-97