Page 34 - 5_TOC in daily practice
P. 34

Sum parameter – Total Organic Carbon

                                         Determination of limit of detection and quantification

                                         for the NPOC method using High Sensitivity Catalyst


                    SCA-130-518

                                                                ■ Calculation of limit of detection and


                                                                quantification according to DIN 32645














            ■ Measurement parameters
            System:       TOC-L CPH
            Catalyst:     High Sensitivity Catalyst
                          (platinum coated wool)
            Method:       NPOC (IC-removal)
            Acidification:  1,5%
            Sparge time:  2 min
            Injection vol.:  408µl                              Characteristics
                                                                Slope a:                           0,031
            ■ Calibration with automatic dilution               Intercept b:                       1,283
            function                                            Correlation coefficient r:         1,000
            A 10-point calibration curve is carried out by      Result uncertainty:                33,3%
            using the automatic dilution of a 1.000µg/L C       Probability of error (a):          5,00%
            stock solution.                                     Number of injections:              3
                                                                Standard error of estimate Sy:     0,078
                                                                Standard error of procedure Sx:    2,513
                                                                © DINTEST

                                                                Limit of detection:           4,180µg/L
                                                                Limit of quantification:    15,352µg/L

                                                                Note:  The result depends on the chosen injection
                                                                volume and the purity of used gases, reagents and
                                                                ultra-pure water.
   29   30   31   32   33   34   35   36