Sum parameter – Total Organic Carbon


                                         Detection and determination limit of TN

                                         method with TOC-L CXN

                    SCA-130-520

                                                                ■ Calculation of detection and


                                                                determination limit according to DIN 32645
                                                                (based on linear regression)














            ■ Measurement parameters
            System:       TOC-L CPN
            Catalyst:     Standard catalyst
                          (Al-balls covered with Pt)
            Method:       TN (with TNM-L)
            Injection vol.:  140µl

            ■ Calibration with automatic dilution
            function
            A 10 point calibration curve is carried out by      Characteristics
            using the automatic  dilution  of  a 2mg/L TN       Slope a:                           30,531
            (based on NO3) stock solution.                      Intercept b:                       -1,847
                                                                Correlation coefficient r:         0,9998
                                                                Result uncertainty:                33,3%
                                                                Probability of error (a):          5,00%
                                                                Number of measurements n:          3
                                                                Standard error of estimate Sy:      0,426
                                                                Standard error of procedure Sx:     0,014
                                                                Sum of squared deviations:          3,3
                                                                Quantile (one-sided):               1,860
                                                                Quantile (two-sided):               2,306



                                                                Limit of detection:         0,023mg/L
                                                                Limit of quantification:    0,084mg/L

                                                                Note:  The results depend on the selected
                                                                       injection  volume, the purity  of  the
                                                                       vessels, water, chemicals and  gases
                                                                       used.