Sum parameter – Total Organic Carbon



                                         TOC –Determination in cleaning

                                         validation - swab method
                    SCA-130-203


            Cleaning     validation  substantiates    the       ■ Sampling and analysis
            effectiveness  of  a  cleaning  process  and        In COP cleaning, the wiping method (swab) is
            ensures that no residues remain on the              used for sampling of visible residues. These
            surfaces of the production equipment. For the       include coatings, crusts, material deposited in
            detection   of    contaminations,   validated       corners and edges, and especially poorly
            analytical  methods must be used that are           soluble  substances.  The swab can be
            sensitive enough to determine the defined           extracted  in  a  solvent  and  the  extracted
            acceptable residue level. In general, residue       solution is subsequently analyzed. If water is
            limits of 10 ppm or  1/1000 of the usual            used for extraction, TOC analysis is suitable
            therapeutic  dose  of  an  active  substance  are   for subsequent analysis. Alternatively, the
            used as acceptance criteria.                        swab can also be measured directly (using a
                                                                carbon-free swab) using a TOC solid-sample
                                                                module.

                                                                ■ Measuring system for the swab test
                                                                The  modular  design  of  Shimadzu’s  TOC-L
                                                                series    now    enables     the   additional
                                                                determination of the swabs using the same
                                                                instrument. For this purpose, a solid-sample
                                                                module (SSM-5000A) was connected to the
                                                                main instrument, either a TOC-L series
                                                                combustion  system  or  the  wet-chemical
                                                                model of the TOC-V series.






            ■ Cleaning methods: Clean out of Place
            For  COP  cleaning,  the  entire  production
            system must be disassembled and the
            components must be  cleaned individually.
            This  procedure  is  very  time  consuming  and
            labor intensive. Due to the individual              For TC determination, the swab is placed in a
            cleaning,   this   procedure     cannot    be       ceramic boat and transferred into the oven,
            standardized. Advantages are, however, the          which is heated to 900 °C.
            low investment costs of the system and the
            possibility of visual inspection.