Page 15 - 2_Pharmaceutical industry
P. 15

Sum parameter – Total Organic Carbon



                                         TOC – Determination according to USP 661.1
                                         Testing of Plastic Packaging Systems and their
                                         Materials of Construction
                    SCA-130-207


            Plastic packaging systems for pharmaceutical        For the TOC determination a purified water
            products must be suitable for  their intended       extraction  is  prepared.  For  example  25g
            use. The US Pharmacopeia revised the                plastic material is  placed in a glass flask,
            related chapter. It is published in USP 39-         500ml purified water added and boil under
            NF34, which will be valid from May 2016.            reflux condition for 5h. After cooling, the
                                                                solution is pass through a sintered-glass filter.
            Besides to the change of the title “Plastic         The parameter of the sample weight, purified
            Packaging  Systems  and  their  materials  of       water volume and heating temperature and
            Construction”,   two   new    chapters    are       time depends of the used plastic sample.
            established.                                        The TOC of the ultra-pure water is subtracted

            This  application  note  is  related  to  the  first   for the measured value of the  extraction
            chapter 661.1.                                      solution. The resultant TOC  value  must not
                                                                exceed 5mg/l.

                                                                ■ TOC determination in pharmaceutical
                                                                application
                                                                The    TOC    determination   is   performed
                                                                according to the USP<643>. This  regulation
                                                                describes  the  TOC  determination  for  pure
                                                                water, purified water and water for injection. It
                                                                does  not  prescribe  any  particular  oxidation
                                                                technique for TOC determination.
                                                                The TOC systems, however, must be able to
                                                                differentiate between inorganic and organic
            ■ 661.1 Plastic Materials of Construction           carbon.  This  can  be  carried  out  either  via
            The purpose of this chapter is to provide test      removal of the inorganic carbon (NPOC
            methods and standards for plastic  materials        method), or via a separate determination
            (e.g polyethylene, polyolefins, polypropylene)      (difference method). The limit of detection for
            of construction used in packaging systems for       TOC should be at least 0.05 mg /L. The
            therapeutic products. The characterization is       applicability of the method must be determined
            done  by  identity,  biocompatibility  (biological   via a system suitability test.
            reactivity),    General      physicochemical        However,  material  extracts  may  have  TOC
            properties  and Additives and extractable           values that are higher than those of purified
            metals.                                             water    because    of   extracted   organic
                                                                substances.
            The TOC  parameter  as an indicator for             Thus the TOC analyses performed should
            extractable   material   is   part   of   the       have a limit of detection of 0,2mg/L and should
            physiochemical  properties  that  must  be          have a demonstrated linear dynamic range
            determined.                                         from 0,2 – 20mg/L.