Page 15 - ICPMS-2040 Series
P. 15

Accessories   Installation






 Sample Introduction System  Installation Information

 Standard Set  For Organic Solvents  For Hydrofluoric Acid Resistance  Temperature within 18 to 28 °C (max. 2 °C/hour change)
 Environmental water, effluent   1. Installation
 Examples of   water, solutions with dissolved   Solution containing       Site Environment  Humidity within 20 to 70 %RH (with no condensation)
 Samples  pharmaceutical or food   Organic solvents  hydrofluoric acid
 substances, or other acid        Avoid using the system in locations with significant vibration or dust.
 decomposition solutions
 Organic Solvent                  Main unit     Single-phase  200 to 240 V ± 10 %, 50/60 Hz,6 kVA
 Mini Torch, 1.5  Mini Torch, HF
 (P/N 211-96077)  Torch  (P/N 211-95846)  Data processing  Single-phase  100 V ± 10 %, 50/60 Hz, 110 VA
 (P/N 211-94021-41)  2. Power Supply
                                  Cooling water   Single-phase  200 to 230 V, 50/60 Hz, 1.2 kVA (standard type)
 Shield Screen,                   circulator
 Shield Screen, Mini  Shield Screen, Mini       Three-phase  200 V, 50/60 Hz, 1.2 kVA (non-CFC type)
 (P/N 211-93819)  Standard  (P/N 211-93819)
 (P/N 211-93820)
 Torch
                3. Grounding      Should be grounded with a maximum resistance of 100  .
 Bonnet, Mini  Bonnet, Standard  Bonnet, Mini
 (P/N 211-95998)  (P/N 211-94047)  (P/N 211-95998)
                                  Type  Argon gas            Min. 99.95 % purity  Pressure: 500 ± 10 kPa
 Torch Adapter,                         Helium gas           Min. 99.999 % purity  Pressure: 200 ± 20 kPa
 Torch Adapter, Mini  Torch Adapter, Mini
 (P/N 211-93779-42)  Standard  (P/N 211-93779-42)  Hydrogen gas  *1  Min. 99.999 % purity  Pressure: 200 ± 20 kPa
 (P/N 211-93780-42)
                4. Gas Supply           Third gas  (gas mixture of 10 % ammonia and 90 % helium)
                                               *2
 Extension Pipe  Extension Pipe, HF                          Min. 99.999 % purity  Pressure: 200 ± 20 kPa
 (P/N 211-95574)  (P/N 211-95847)       Oxygen-argon mixture gas  (70 % argon and 30 % oxygen)
                                                            *3
 Extension Pipe                                              Pressure: 450 ± 10 kPa
 Dilution Gas Tube
 (P/N 211-95989)
                5. Cooling Water  Use a dedicated cooling water circulator.
 Cyclone Chamber  Cyclone Chamber, HF                     3
 (P/N 211-96078)  (P/N 211-95848)  6. Exhaust Duct  Exhaust flowrate of 2.4 to 3.3 m /min at exhaust outlet (install a damper).
 Chamber
 Helix CT Locking Screw with Seal  7. Dimensions  W853 × D660 × H554 mm
                                  (excluding the plasma stand exhaust duct and other protrusions)
 (P/N 046-00093-95)
                8. Weight         144 kg
 Nebulizer, DC04  Nebulizer, HF
 (P/N 211-95988)  (P/N 211-95845)  Note: Refer to the pre-installation requirements for more details.
 Nebulizer     *1: ICPMS-2050 Series systems only
 Carrier Gas Tube  Suction Tube, NFT-050  *2: When ICPMS-2050 Series systems are equipped with an optional third-gas system
 (P/N 046-00092-53)   (P/N 211-97567-41)  *3: When an optional organic solvent injection system is installed
               For more detailed system specifications, refer to the specifications sheet.
 Drain  Drain Trap  Drain Trap, HF
 (P/N 211-93814-01)  (P/N 046-00093-06)
 Pump Tube, 0.76-BLK3B-95-SF-F    Example of External Dimensions and
  (P/N 018-31558-61)  *1
 Peristaltic   Pump Tube, 0.76-BLK3B-95-F  Pump Tube, 0.64-OW-MH-95-F    Pump Tube, 0.76-BLK3B-95-F  Installation Configuration
 Pump Tube  (P/N 018-31558-44)  (P/N 018-31558-32) *2  (P/N 018-31558-44)
 Pump Tube, PUR-T,381-2232F
 (P/N 016-46043-01) *3
                                              Helium gas cylinder                                unit: mm
 *1: For ethanol/methanol/IPA   *2: For DMF/NMP   *3: For xylene/kerosene  497  Hydrogen gas cylinder


 Interface      Min. 500  296                                     554
                                                                                       300
 Standard Set  Option
 Nickel Sampling Cone  Platinum Sampling Cone  660
 Sampling Cone  -
 (P/N 211-97283-03)  (P/N 211-97283-04)
                                                               615
 Nickel Skimmer Cone, Small                                                       295
 (P/N 211-90200-43)  Platinum Skimmer Cone  Copper Skimmer Cone
 Skimmer Cone
 Used in combination with spacer   (P/N 211-90194-02)  (P/N 211-90200-41)  377  853  320  600
 for small skimmer (P/N 211-95342-01).
                                    1400

 14                                                                                                               15