Page 66 - Pharmaceutical Solution for Pharma Analysis
P. 66

Analysis of styrene leached from polystyrene cups
       using GCMS coupled with Headspace (HS) sampler






                                                 Table 1. Method validation parameters
                                           Parameter                  Concentration (ppb)
                                           Linearity                   1, 2.5, 5, 10, 20, 50
                                       Accuracy / Recovery                2.5, 10, 50
                                       Precision at LOQ level                1
                                         Reproducibility                     50





          HS-GCMS Analytical Conditions
          Figure 3 shows the analytical instrument, HS-20 coupled with GCMS-QP2010 Ultra on which samples were analyzed with
          following instrument parameter.





















                                        Figure 3. HS-20 coupled with GCMS-QP2010 Ultra by Shimadzu




          HS-GCMS analytical parameters

                        Headspace parameters

                          • Sampling Mode     : Loop
                          • Oven Temp.        : 80.0 ºC
                          • Sample Line Temp.    : 130.0 ºC
                          • Transfer Line Temp.    : 140.0 ºC
                          • Equilibrating Time    : 20.00 min
                          • Pressurizing Time    : 0.50 min
                          • Pressure Equilib. Time   : 0.10 min
                          • Load Time         : 0.50 min
                          • Load Equilib. Time    : 0.10 min
                          • Injection Time    : 1.00 min
                          • Needle Flush Time    : 10.00 min
                          • GC Cycle Time     : 23.00 min






                                                                                                                      3