Drug Discovery



            Protocol                                           Analytical method

            Incubation conditions                              A LC-MS/MS method is developed for each compound. In addition, a
            • Test Compound concentration: 10 µM               calibration curve is established to quantify the amount of compound
            • Microsomal Protein concentration: 0.5 mg/mL      during metabolism. Two approaches are used.
            • Incubation time: 60 min                          • Disappearance of the parent ion mass over a period of time
            • Co-factors added: NADPH                          • Accumulation of the metabolites over a period of time
            • Internal standard: Verapamil
                                                               Different SCAN and SIM modes of the quadrupoles useful in the analysis
            • Positive control: HZ-166
                                                               • Q3 Scan
            • Negative control: without microsomes
                                                               • Selective Ion Monitoring
            • Time points: 0,10,20,30,40,50 and 60 min
                                                               • Product ion scan
            • Instrument: Digital Dry bath
                                                               • MRM Transition
            • Temperature: 37°C
                                    Microsomal Assay mixture
                                    incubated at 37°C
                                                                                  Compound      Internal
                                                                                                standard
                                                                    Intensity
                                    Ice cold Acetonitrile
                                    with Verapamil



                                    Sonication followed by
                                    centrifugation                                 Retention time



                                    Centrifuge filter tubes




                       Analysis by LCMS 8040