Page 11 - Shimadzu inspeXioSMX-225CT FPD HRPlus
P. 11
Optional Module
Optional Module “CORE Boost ”
™
“CORE Boost” is an optional module that extends the 2 functions, “Boost Reconstruction” and “ROI
Reconstruction”, to the control software for the inspeXio SMX-225 CT FPD HR, inspecXio64.
Boost Reconstruction
Boost Reconstruction Function is an image reconstruction calculation process that uses our proprietary
high-speed calculation technology for image reconstruction calculation processing of high-pixel CT
images to achieve even higher speed.
Faster image reconstruction at 2048 × 2048 pixels or 4096 × 4096 pixels.
About 20 times faster
60 seconds
New (Boost Data acquisition or less Display
Reconstruction Data
Enabled) Calculation
Save
Past (Boost Data acquisition About 1200 seconds
Reconstruction
Disabled) Calculation Save Display
Data
* The above shooting conditions are as follows: clear mode, full scan, 2400 View, 1500 slices, image reconstruction size 2048 × 2048.
* Since Boost Reconstruction Function is a speed-specific image computation technology, it may affect the quality of cross-section images.
ROI Reconstruction
ROI Reconstruction Function is an image reconstruction calculation process that uses our unique ROI
reconstruction technology to improve the image quality of data taken when a part of a workpiece
protrudes (truncates) from the left and right sides of an X-ray uoroscopic image during CT imaging.
This reduces shading artifacts that traditionally occur from the truncation to the entire image,
making it easier to analyze with image processing software.
What’s “Truncation”…? Image Comparison Advantage
A part of a workpiece protrudes ROI image reconstruction The reduced shading artifacts
from the left and right sides of a reduces shading artifacts more make thresholding easier.
uoroscopic image during a CT than traditional methods, so
scan. grayscale values are close to true
value and uniform.
FOV
Line Pro le
3
X-ray
Grayscale Value 2.5 New Method
Detector 2 True value: 2.7 Traditional Method
1.5
1
0.5 Traditional Method
True Value
0
Reduce
shading
artifacts.
* Comparison of conventional and new line profiles.
Transected CT Image ROI Reconstruction
11