Observational Examples of Hydration/Solvation Structure




 What are Hydration and Solvation?  The Hydration/Solvation Measurement Method

 It is known that liquids in contact with solids become stratified, a   1. Run the HR-SPM in a solution, bringing the cantilever close to the surface of the solid sample with high precision, up to the settings value (fmax).
 Water molecule
 phenomenon called solvation, or in the case of water, hydration.   2. Use the force curve method to measure the force on the probe on the tip of the cantilever.
 It is believed that this characteristic structure, which is different   Hydration  3. Variations in characteristic power ( f) in accordance with the sample can be obtained in the immediate vicinity of the solid-liquid interface.
 distance
 from that of a bulk liquid, is the main influence on the various   4. Variation of the force is due to hydration/solvation, which provides insight into the stratification structure in the liquid.
 r23
 roles played by solid-liquid interfaces, such as dissolution,   5. A cross section of the hydration/solvation structure can be visualized by performing continuous acquisition in the X-direction (Z-X measurement).
 r12
 chemical reactions, charge transfers, wetting, lubrication, and   Solid-liquid  6. Furthermore, the 3D structure of the solid-liquid interface can be analyzed by acquiring Z-X measurements in the Y direction.
 heat transfer in the liquid phase. However, because this layer is   Solid sample  interface  Tip Scanning Trace
                            z              z                           z
 extremely thin, the hydration/solvation structure is not easy to   AFM Tip     Water   AFM Probe
 Model of Hydration Structure                                                  molecule
 measure experimentally. In particular, non-uniform structures in              Density
                                                                                          Large      Interaction force
 the in-plane direction of the surface have not yet been detected.  Measurement  Sparse          Small
                                                        Z-X
                                     Variation                                    Dense                 Large
                                       of
 Differences from Existing AFM Technology  the Force                             Sparse
                                0      f                                          Dense
 When measuring hydration/solvation, the variation in the force on the cantilever is very small. However, by using an    f max
                              Force curve                       x                 Dense
 ultra-high sensitivity FM method, this can now be measured for the first time.  y
                                                                      Density distribution function
 Such methods enable not only the measurement of hydration/solvation structures in the Z-X direction, but also the   g (z)
                                            Sample surface
 analysis of 3D Z-XY structures. HR-SPM capabilities have now advanced from mere surface observations to   Pattern Diagram of Hydration/Solvation Measurement  Diagram of Hydration/Solvation Measurement
 measurements of the structure of solid-liquid interfaces.

 Force Curve by Existing AFM  Force Curve by FM Method  Data Analysis Software

                    This software is designed specifically for 3D mapping data analysis. It provides powerful support for analyzing hydration
                    and solvation structure data.








                             2D image data                                                           3D display

 Hydration Measurement Using the FM Method  4)  Observation and Analysis of 3D Hydration
 Structures at a Solid-Liquid Interface  5)
 Z

 1.5 nm  3rd layer
 2nd layer
 1st layer
 Z=1.6nm  Z=1.6nm
 Surface                       1D data
 Z-X mapping of  f  3 nm  X
 The layered hydration structure  X=3nm  Y=2.3nm  X=3nm  Y=2.3nm  Displaying 3D mapping data
 of a mica surface can be visual            Extracting and displaying 2D images from the mapping data
 All data on this page was obtained using the petri dish type solution cell (option).  Displaying and analyzing specified 1D data over 2D image data