Page 20 - Shimadzu RF-6000
P. 20
Extensive Selection of Accessories
Integrating Sphere Unit
(P/N 207-21460-41)
Used in combination with the dedicated application function for
LabSolutions RF to determine the fluorescence quantum yield and
fluorescence quantum efficiency.
Samples : Liquids, solids, or powders
Internal diameter of integrating sphere : 100 mm
Integrating sphere material : Spectralon
Max. sample size : W12.5 × H45 × T12.5 mm
Measurement wavelength range : 200 to 900 nm
Standard contents
・ Integrating sphere attachment (main unit)
・ Mesh for measurement of Spectral correction functions
Ultra Micro Cell Holder Unit
(P/N 207-21455-41)
This ultra micro cell holder unit allows sample quantities below 100 µL
to be measured. It uses a commercial micro cell.
Standard contents
・ Ultra micro cell holder
Available micro cell
• Hellma Cat. No. 105.250-QS-15 (Min. sample amount 100 µL)
• Hellma Cat. No. 105.251-QS-15 (Min. sample amount 45 µL)
Mercury Lamp Unit
(P/N 207-21700-41)
Used to inspect resolution, wavelength accuracy, and wavelength
repeatability for validation.
Standard contents
・ Mercury lamp (main unit)
・ Scattering block
Constant-Temperature Single Cell Holder with Stirrer
( P/N 206-24930-41 (100 V)
206-24930-58 (220/240 V))
206-24930-42 (115 V)
Used to stir and measure temperature-controlled samples. This is ideal for measuring the fluorescence of
suspended cells. Enables controlling the temperature of one cell by circulating temperature-controlled
water. The operating temperature range is 5 °C to 70 °C. The stirrer rpm is variable. The minimum sample
volume required is 2.5 mL.
For use with 100 V, 115 V, or 220/240 V power supplies
Note 1: This cannot be used if there is a large difference in specific weight between the sample and solvent.
Temperature-controlled water tubing connectors are compatible with 4 to 8 mm ID tubing.
Note 2: Requires a separate optional front cover (P/N 207-20490).
20