Page 9 - Shimadzu PDA-7000
P. 9

PDA Time-resolution Analysis






 Shimadzu's unique PDA photometry is installed as standard to offer a variety of benefits for various analyses.
            Reduced Effects of Sample Defects (cracks, pinholes) on Analysis Values
 PDA Photometry (Pulse Distribution Analysis Photometry)  (Patented)

 Unlike conventional photometry that simply integrates the   determine the percentage content.  The internal standard monitoring method is applied to each   To enhance reproducibility,
 photocurrent obtained from discharges over a fixed time, the   PDA permits speciation analysis and a variety of data   discharge pulse to monitor the light intensity for internal   statistical processing is used to
 PDA method integrates the photocurrent obtained from each   processing not possible with previous methods, which   standard elements. If this value lies within the specified range,   eliminate the data obtained
 discharge pulse and analyzes each integral value as having an   significantly increases the analysis accuracy for many elements.   the light intensities obtained for the measured elements are   when the spark is applied to a
 individual significance. The optimal processing is then used to   selected.  defect resulting from sampling.
 Time-resolution Analysis (Patented)

 The discharge conditions to obtain the optimal measurement sensitivity differ for
 Time-resolution Photometry
 each spectral line.
 I                                               Monitoring with Internal Standards
 The diagram at the right shows the current waveform for typical discharge
 conditions. Some elements exhibit sensitivity in Area A, others in Area B. Analyzing
 the elements separately in the individual areas dramatically enhances the sensitivity   Spectral intensity
 for trace elements.   A  B
 The table below compares the Background Equivalent Concentration (BEC) values for   T
 a conventional method and the PDA-7000.   Spectral intensity  Spectral intensity(   )  Spectral intensity(   )
 The lower the BEC value, the higher the sensitivity for trace elements.

 Element  C  P  S  B
 T                                           0
 Conventional method(ppm)  160  150  100  80
 Time             Frequency of occurrence         Internal standard elements Time (T)  Measured elements  Time (T)
 PDA-7000(ppm)  80  75  50  40
                                                          : median value of internal standard elements  ɺ  : discriminator levels
 Shorter Analysis Times (just over 10 seconds per analysis)
 Combined discharge and PDA time-resolution photometry obtain results in just over ten seconds per analysis.

 Data is Controllable for each Discharge, Allowing Photometric Values for
 Abnormal Discharge Pulses to be Eliminated for Enhanced Accuracy  Acid-soluble Speciation Analysis (Quantitation of Acid-soluble Aluminum)
            Statistical processing based on the differences in light intensity between acid-soluble aluminum (sol-Al) and
 Conventional photometry determines the overall integral value   and then conducts statistical processing, thereby achieving
            acid-insoluble aluminum (insol-Al) allows quantitation of acid-soluble aluminum in just 1 seconds.
 of the light intensities. However, the PDA-7000 Series   measurements that are unaffected by light intensities from
 determines the frequency distribution of the light intensities   abnormal discharges.  Analyzed Sample Surface  Principle of Speciation Analysis of Aluminum in Steel

 Enhanced Accuraacy for Insoluble Elements that Readily Form Inclusions   m : median value of analysis values
                           Insol AI
 (Patented)                                 Discharge including insol-Al  N : total number of pulses  sol AI
                                                                          sol AI     m×  n    Measured value
                                                     Discharge for sol-Al only  n samples  N
 Achieves two or three times the reproducibility of conventional   conventional method and the PDA-7000 for elements that
 Shimadzu methods.   readily form inclusions.                                     insol AI     Chemical Analysis of sol-Al
                                                                     Frequency of occurrence
 The table at the right compares reproducibility values for a                        n’ samples  insol AI
 Element  Total Al  S  Pb  B  Ca                                                       n      Measured value
 Content (%)  0.037  0.022  0.005  0.0035  0.002                                     m×  N
 Reproducibility   Conventional method  0.0025  0.0010  0.0010  0.00027  0.00045  Discharge spot  m  Spectral intensity  Chemical Analysis of insol-Al
 σ (%)  PDA-7000  0.00045  0.0004  0.0002  0.00005  0.0001