Page 25 - Shimadzu IRTracer-100
P. 25
MCT Kit (P/N 206-36050-58)
Use a high-sensitivity MCT detector for analyses where a large between the standard DLATGS detector and the MCT detector is
amount of light is not available, such as monomolecular lm performed automatically from LabSolutions IR. In addition, the
analysis on metal substrates, high-speed reaction tracking, and software has a built-in liquid nitrogen sensor to terminate current
low-concentration gas analysis using a long path-length gas cell. ow when the detector element is not being cooled, thus protecting
The kit installs an MCT detector on the IRTracer-100. Switching the MCT detector.
Note: This kit cannot be mounted at the same time as the Near IR Kit (P/N 206-74253-58). Liquid nitrogen is required when using the MCT detector.
Far IR Kit (P/N 206-30616-58)
Bands related to inorganic compounds and organometallic complexes are 3
−1
typically observed in the far infrared region to 240 cm . The Far Infrared Kit Abs
contains a Csl beam splitter that can be mounted on the IRTracer-100 for 2.5
measuring spectra in this region. 2
−1
−1
Since absorption due to water is greater in the 400 cm to 240 cm , the
instrument should be purged with desiccated air before performing 1.5
measurements. The Csl beam splitter should also be stored in a desiccator 1
when not in use since it is highly sensitive to moisture.
0.5
0
2000 1800 1600 1400 1200 1000 800 600 400
Spectrum of Polyvinylidenechloride PVDC 1/cm
Near IR Kit (P/N 206-74253-58)
Attached to the IRTracer-100, this kit enables near-infrared Specifications Measurement Range: 12,500 to 3,800cm −1
measurement. LabSolutions IR switches between the mid-infrared Beam Splitter: Si/CaF 2
and the near-infrared. Light Source: Tungsten iodine lamp
Note: This kit cannot be mounted at the same time as the MCT Kit (P/N 206-36050-58). Detector: InGaAs
NIR Integrating Sphere IntegratIR A (P/N 208-97272-92)
Powders, tablets, liquids, bers, plastic pellets and molded samples can be
placed on the sample stage for measurement (re ectance measurement).
Pre-treatment such as KBr dilution is not required.
Samples stored in a plastic bag or glass bottle can be measured.
Applications include qualitative or identi cation tests in acceptance
inspections and quantitative analysis of components in measured samples.
Features a built-in highly sensitive InGaAs detector.
Note: The IntegratIR installation kit (P/N 206-72715-93) must be purchased separately.
Please contact your Shimadzu representative about accessories that do not appear in this brochure.
For Other Accessories
Also note that it may not be possible to use FTIR-8000 Series accessories. Please consult your Shimadzu representative for assistance with using FTIR-8000 Series accessories.