EZ Test Speci cations  Options





 EZ Test     Tester Options
 Name
 EZ-SX  EZ-LX  EZ-LX HS
 Max. 500 N  Max. 5 kN  Max. 2 kN  Jog Controller                                   Safety Cover
                                 346-55922-01                                       EZ-SX: 346-57107-01
 Tester Load Capacity  (note 1)  The load cell type can be selected from 9 types:   The load cell type can be selected from 12 types: 1 N, 2 N, 5 N, 10 N,   The jog dial is provided to allow  nger-tip   EZ-LX: 346-57107-02
 1 N, 2 N, 5 N, 10 N, 20 N, 50 N, 100 N, 200 N, or 500 N.  20 N, 50 N, 100 N, 200 N, 500 N, 1 kN, 2 kN or 5 kN. (Up to 2 kN for EZ-LX HS)  operation of the crosshead position.  This is used to prevent injury when sample
                                                                                    fragments are scattered due to fracture.
 Load Method  High-precision constant-speed strain measurement using backlash-free ball screw drive
 ±0.5% of indicated value (within 1/500 to 1/1 of load cell rated capacity)
 High-Precision Type  (note 2)
 Conforms to JIS B 7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4.
 ±1% of indicated value (within 1/500 to 1/1 of load cell rated capacity)  Control I/O Expansion Box  Analog Recorder
 Test Force Measurement  Standard-Precision Type  (note 2)  346-55920-01
 Conforms to JIS B 7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4.  Increases the number of control I/O ports to   X-T type: 346-59210-01
 Range  1 range (rangeless)      four. Multiple optional accessories can be         Plots test force - time curves.
 Test Force Calibration  Automatic calibration using calibration cable  simultaneously connected to the control I/O   X-YT type: 346-51736-01
                                 ports.                                             Plots test force – time curves and test force –
 Crosshead Speed Range  0.001 to 1000 mm/min  0.001 to 2000 mm/min
                                                                                    stroke curves.
 Maximum Return Speed  1500 mm/min  3000 mm/min
 Crosshead Speed Accuracy  (Note 3)  Within ±0.1% of test speed
 Crosshead Speed and Allowable Test Force  Up to the capacity of the load cell at all speeds  Sensor I/O Expansion Box  Power Cable
                                 346-55920-02                                       For EU (VDE standard)
 Distance Between Crosshead and Jig Mounting Surface  500 mm  920 mm  Increases the number of tester sensor I/O   348-34063-03
 700 mm (5 kN load cell + 5 kN screw type  at grips)  ports to two. Multiple optional accessories
 Maximum Grip Space  395 mm (500 N max. load cell + tensile jig)  755 mm (1 kN load cell + 1 kN screw type  at grips)  can be simultaneously connected to the   For China (GB standard)
                                                                                    348-34063-02
                                 sensor I/O ports. BNC cables can be
 860 mm (500 N max. load cell + tensile jig)  connected to the analog I/O ports.    For Japan and North America (UL, CSA, PSE standards)
 Depth of Test Space  100 mm (table section)                                        348-34063-01 is provided as standard.
 Crosshead Position  Measurement & Display  Optical encoder measurement, digital display (display resolution: 1 µm)  A variety of other options are also available. For details, refer to the separate catalog "Optional Accessories for Autograph".
 Detection  Accuracy  ±0.1% of indicated value or ±0.01 mm, whichever is greater
 Crosshead Control  Single test control (single-direction tension or compression test), cycle test control (repetitive tension or compression test)
 Sampling Speed  1 ms max. (TRAPEZIUM X/TRAPEZIUM LITE X is needed for this function)
 Constant test force (creep) control  (note 4)  Additional Load Cell Kits
 Auto-stop and auto-return functions when specimen fracture is detected (crosshead auto home-position return)
 Test condition  le function, adjustable crosshead speed function  Select a load cell kit if load cells are to be added to the tester unit kit. The additional load cell kit comprises a cell set (load cell and
 Display function: Actual test force display or stress display (set by user)  calibration cable), cell bolt (if required), and upper joint jig (if required).
 Crosshead displacement display in mm or %/GL (set by user)
 Standard Functions Included  Peak point test force and stroke  LOAD CELL SET (Load cell, calibration cable and calibration)
 Test force and displacement analog output: 0 V to 5 V DC full scale, respectively (for external recorder)
 USB interface                                                                 EZ-LX
 Manual crosshead position  ne adjustment  CLASS  EZ-TEST  -                     EZ-LX HS
 Adjustable controller                              -                                EZ-SX
 Touch load alarm               P/N          5 kN  2 kN  1 kN  500 N  200 N  100 N  50 N  20 N  10 N  5 N  2 N  1 N
 Dimensions and Weight  W400 × D530 × H885 mm, approx. 33 kg  W400 × D530 × H1315 mm, approx. 55 kg  1  346-55939-XX  10  14  9  13  12  07  06  05  04  03  02  01
 Input Power Supply Voltage  (Note 5)  Single phase, 100 V to 150 V AC, 50/60 Hz, or 200V to 230V AC, 50/60 Hz  0.5  346-55942-XX  10  14  9  13  12  07  06  05  04  03  02  01
 Power Capacity  700 VA  850 VA
 Temperature: 5°C to 40°C, Humidity: 20% to 80% (no condensation)
 Installation Environmental Conditions
 Power voltage  uctuation: within ±10%, Vibration: 10 Hz max., Amplitude: 5  m max.
 Note 1: When the load cell capacity is smaller than the tester load capacity, the former is the maximum test force.   Thermostatic Chamber  Installation Space
 Note 2: Shimadzu recommends inspection at an installation site that meets the requirements speci ed in these standards.
 Note 3: Crosshead speed accuracy is calculated from the crosshead displacement within a prescribed time at a constant speed between 0.5 mm/min and 500 mm/min.  Allows testing within an ambient
 Note 4: The test force is kept constant at 70% or less of the tester load capacity, for up to 12 hours.   (unit: mm)
 Note 5: Ground resistance should be 100   or less.  temperature range of -70°C to 250°C.   Wall surface
            Thermostatic chambers are available only
            for EZ-L type testers.                                                 Power supply input
             The chambers are special-order items. For                 Anti-topple attachment   (710 mm above  oor)  600  Worktable
                                                                                   (When a worktable of
                                                                       (included in the tester
            details, please make a separate inquiry.                   unit kit)   height 650 mm is used)  (optional)



             Other Options                                                    250
                Description      P/N              Remarks                     125
                Worktable     340-48580-02     Worktable for EZ Test             38
                                                                            600        400       485
                              346-55037-12  Secures the tester unit to the worktable.
             Anti-topple attachment                                                    420       475
               for EZ-SX/EZ-LX  346-55037-11  Secures the tester unit to the worktable,
                                              or the worktable to the  oor.
                                                               Note: A similar installation space is required for both EZ-SX and EZ-LX.