Automatic  Automatic
                                                                                        Contaminant  Qualitative
                                                                                        Recognition  Analysis
                                                                               Wide-Field                      High
                                                                                Camera   Finally, a wide view   Sensitivity
                                                                                         on micro sample analysis
 Accessories
                                                                                  Contaminant             Micro
                                                                                    Analysis   Surface    Analysis
                                                                                               Analysis






 ATR Objective (slide-on type)   Ge prism: P/N 206-32600-41  DLATGS Detector   P/N 206-32580-41
 The ATR objective uses a cone-type prism, with single reflection, 15× magnification and a 45-degree incident angle. The   By adding this DLATGS detector to an AIM-9000 infrared microscope,
                                                                         DLATGS detector
 slide-on type prism makes it easy to switch back and forth between visible observation and infrared measurement. This ATR   infrared spectra can be obtained without liquid nitrogen. It is also   MCT detector
 objective is especially effective in analyzing samples that do not transmit or reflect infrared light easily, such as paper and   possible to switch back and forth between the MCT and DLATGS
 plastics, or extremely thin film, such as stains.  detectors for measurements as needed. The DLATGS detector offers a
                                                                   Abs
                                          -1
               wider wavenumber range (up to 400 cm ) than the MCT detector, but
 Accretion     with lower sensitivity. Therefore, the MCT detector is used to measure
 Fiber
               micro samples less than 100 µm.
                                                                    4000  3000  2000  1500  1000   500 cm -1
 Abs                                                                   The infrared spectra of polyvinyl chloride (PVC)

               Diamond Cell CII   P/N 208-92289-01

 4000  3000  2000  1500  1000  cm -1  This diamond compression cell is used to
               compress micro samples very thin for
 Spectra of contaminants on a textile identified as Phenolic resin
               direct measurement under the
               microscope. It can be used for samples                      After rolling
               such as plastics and fibers. This CII cell                  Before rolling
 ATR Pressure Sensor   P/N 206-32603-41  features a large thin window plate made
               of artificial diamond (1.6 mm diameter).              Abs
 This pressure sensor prevents prism damage due to excessive pressures applied
               A type-B cell that uses natural diamond is
 during ATR measurements using an ATR objective. It can also be used to
               also available.
 automate ATR measurement with pressure sensing.

                                                                      4000   3000   2000   1500  1000 cm -1
                                                                            The infrared spectra of a single fiber


 Grazing Angle Objective (GAO)   P/N 206-32602-41  Micro Vise Holder   P/N 206-33293
 The Grazing angle objective with 80-degree incident angle is effective for measurement of organic thin film with nm level on   This holds various types of samples for microscopy. It ensures positive holding of samples of a difficult shape or measurement of a
 metal substrate. In the case of failure analysis, this objective is useful for measurement of samples on concave surfaces or   sample at a user-selectable angle. Measurement with a polarizer, with the sample under tensile load, provides information on the
 stains on a metal surface.  molecular orientation.

 Microscope image using the visible polarizer
 Grazing angle objective (GAO)  Infrared Polarizer   P/N 206-32605-41
 Cassegrain objective
               This accessory is useful for researching the orientation characteristics of samples, or increasing sensitivity with Grazing Angle
               Objective measurements. The infrared polarizer can be used by inserting it into the microscope from the side.

 Abs
               Visible Polarizer   P/N 206-32540-41
               This accessory is useful for visible observation of samples that are normally difficult to observe using visible light. Using the
 4000  3000  2000  1500  1000 cm -1  properties of polarized light can make samples easier to see.
 The infrared spectra of stains on the metal (oil film)



 8                                                                                                                 9