World-Class High-Sensitivity Furnace




                                                                                                                                      High-Performance Autosampler

                                                                                                                                         Mix and inject up to four sample types (diluted solution, standard solution,
                                                                                                                                         sample, matrix modifier, etc.). (Unmixed samples can also be injected.)
                                                                                                                                         Select a fluororesin tube or pipette tip as the injection nozzle.
                                                                                                                                         Automatic dilution and re-analysis if the calibration curve concentration is exceeded.



                                                                                                                                      Example of Automatic Dilution and Re-Analysis
                                                                                                                                      Analysis of cadmium (Cd): The maximum concentration of the calibration curve has been set to 1ppb. At 1.8ppb, the sample of [003] has exceeded the
                                                                                                                                      maximum concentration. As a result, it has been automatically diluted by four times and re-measured.
















                                                                                                                                      High-Sensitivity Application Data (Analysis of Selenium (Se))

                                                                                                                                         Independent control of the gas flow through the graphite tube during
                                                                                                                                         atomization achieves extremely high-sensitivity measurements.
                                                                                                                                         (Japanese Patent 2067563)





                                                                                                                                                       Measurement Example Showing High Sensitivity
                                                                                                                                                       Analysis of 0 to 5 ppb selenium (Se): Sensitivity is adequate for 1 ppb measurements.
                                                                                                                                                       (20 µL injected volume, Pd modifier, pyro-coated graphite tube)


                                                                                                                                         Digital temperature control and electronic gas flow control enhance stability
                                                                                                                                         Combination of a high-sensitivity optical sensor and unique digital temperature control
                                                                                                                                         technology achieves highly accurate temperature control in all temperature regions from
                                                                                                                                         drying to atomization. An electronic flow controller can precisely control the inner gas
                                                                                                                                         flow rate in 0.01 L/min units. These control techniques significantly enhance both the
                                                                                                                                         sensitivity and the data stability.

                                                                                                                                                       Measurement Example Showing High Stability
                                                                                                                                                       Analysis of 0.5 ppb lead (Pb): Approx. 5% relative standard deviation after five
                                                                                                                                                       repeated measurements is adequate for quantitative analysis.










            Achieves World-Class Lower Limits of Detection
            Advances in optics and a new graphite furnace design achieve improved   Enhanced Sensitivity due to Graphite Furnace
            lower limits of detection for furnace analysis (comparison with previous   Previous       New
            Shimadzu models).                                                       Pb 0.08ppb       0.05ppb
            This superb analysis performance is possible in all fields.             Mn 0.02ppb       0.01ppb