Page 56 - Application Handbook - TOC
P. 56

SCA-130-502



                                                                POC-Peak:















                                                                NPOC-Peak:




                  Fig. POC-Calibration with IC-standard


            ■ System testing
            In POC analysis, it is of great importance that
            the LiOH trap functions perfectly. A test
            should, therefore, be carried out to confirm
            the efficiency of the CO 2 trap.  This test
            should be carried out each working day:             Result:
            An IC control solution (TIC = 1000 mg/L) is
            prepared and analyzed as a POC sample.              TOC=4,05mg/l
            For efficient functioning of the LiOH trap, the     POC=4,05mg/l   =>  100%
            POC measuring result must be < 0.1 mg/L.
                                                                NPOC=0mg/l

            ■ Example of a POC measurement

            Sample: Toluene in ultrapure water                  ■ Recommended Analyzer / Configuration
                                                                TOC-L CXX
            Toluene is a compound that is  fully                ASI-L
            purgeable. In the  addition method, the             POC-Option
            toluene sample is completely determined as
            POC.

            POC measuring method:
            Injection volume:     800 µL
            Purging time:         3 min
   51   52   53   54   55   56   57   58   59   60   61