Page 13 - Application Handbook - TOC
P. 13

SCA-130-105



            The suspension is homogenized for several           ■ Example of suspension determination
            minutes at a speed of 17,000 – 18,000 rpm
            using a dispersion unit. It is critically
            important to select a suitable precision tool.
            After homogenization, the suspension is
            transferred directly to autosampler vials.
            Critical for  subsequent measurement is the
            use of a magnetic stirrer in the autosampler
            to ensure that the suspension remains
            homogeneous during sampling. After all, the
            small particles should not sediment, but
            remain uniformly distributed throughout the
            solution. The suspensions can now be
            repeatedly injected and analyzed.                   NPOC-Method (Acidification is done by
                                                                creation of the suspension)
                                                                Sparge-time: 5 Minutes
                                                                Injection volume: 90µl
                                                                Multi-Injection is deactivated

                                                                ■ Statistic

                                                                Peak-No     Areas      Result of       Result of
                                                                                      suspension     Sample TOC
                                                                                      TOC in mg/l        in %
                                                                    1       183,7        24,71           2,47
                                                                    2       180,9        24,23           2,42
                                                                    3       189,0        25,61           2,56
                                                                    4       183,8        24,72           2,47
                                                                    5       179,9        24,06           2,41
                                                                    6       179,9        24,06           2,41
                                                                    7       179,9        23,89           2,39
                                                                    8       186,9        25,25           2,53
                                                                    9       181,6        24,35           2,37
                                                                   10       177,6        23,67           2,37
                                                                   MW       182,2         24,5           2,45
            This method enables  not only automated             RSD in%      1,88         2,52           2,49
            solid sample analysis using an autosampler,
            but also parallel measurement of solid and
            liquid samples in the same sample table and         ■Recommended analyzer / Configuration
            the same sample tray. In addition to speed,         TOC-L  CPN
            this method is also  impressive in terms of         ASI-L (40ml) with stirrer option and
            robustness. The possibility of multiple             external Sparge-Kit.
            injections  enables differentiation of small
            variations due to weighing and measuring.
   8   9   10   11   12   13   14   15   16   17   18