Sum parameter – Total Organic Carbon




                                         Carbon dioxide determination in beer

                   SCA-130-403


            Carbon  dioxide  is  an  important  ingredient  in   In order to develop a method that does not
            many soft drinks. This is also the case for         have these disadvantages, a TOC analyzer
            beer. It creates a sparkling and refreshing         was used.
            (tangy) taste and is important for the
            formation of foam.                                  ■ Innovative methods
                                                                In  this  method,  the  sample  (beer)  is  directly
            The CO 2 content of a beer affects the              placed in a 40 mL autosampler vial. 5 mL of a
            threshold  values  for  various  fragrance  and     32%  NaOH  solution  was  added  to  the
            aroma  components.  In  addition,  bottling         autosampler vial to preserve the CO 2.
            under CO 2 increases the shelf life of beer..
                                                                The sample is subsequently added directly to
                                                                the autosampler and the IC  (inorganic
                                                                carbon) content is measured.





















                                                                Preservation step:
            In the manual of the ‘central- European
            brewery  technological  analysis  commission)’
            (MEBAK)      various    methods     for   the                            -             -
            determination of CO 2 are  listed.  These are                  CO  + OH     →   HCO
                                                                                                  3
                                                                              2
            generally  based  on  manometric  or  titrimetric
            method, or they are methods that use                                   -           2-
            specialized detectors.                                    CO  + 2 OH      →  CO  + H O
                                                                         2                    3      2
            Disadvantages  of  these  methods  are  often
            the lack of selectivity for CO 2 (other gases or    In the TOC analyzer, the sample is injected in
            substances are also determined), high               a concentrated phosphoric acid solution
            expenditure in terms of personnel and time,         (25%). The CO 2 is subsequently released
            and the lack of possibilities for automation.       again and is transferred via the carrier gas to
                                                                a CO 2-selective NIDR detector where it is
                                                                detected.