Page 20 - 3_Chemical industry
P. 20

SCA-130-309



            Automated reagent preparation eliminates            The autosampler OCT-1 has a big advantage
            any  contamination  of  the  reagent  solutions     for the measurement of aggressive matrices,
            and  minimizes  the  blank  value  of  the          because its parts are done of Teflon.
            instrument.
                                                                ■     Method      of    phosphoric      acid
            ■ TOC determination                                 measurement
            Due to lower TOC concentration, the                 The concentrated phosphoric acid (85%) is
            instrument is calibrated in a range of 0,1 – 1      diluted by 1:5.
            mg/L.                                               TOC  determination  in  phosphoric  acid  was
                                                                carried out using the NPOC (Non-Purgeable
                                                                Organic Carbon) method. Prior to TOC
                                                                determination, neutral or alkaline samples are
                                                                acidified in order to decompose all inorganic
                                                                carbonates and bicarbonates. This step could
                                                                be omitted for phosphoric acid.

                                                                Acidification:       0%
                                                                Sparge-time:         2 minutes
               Figure: Calibration of NPOC between 0,1 – 1 mg/l.   Injection volume:    3000µl

            Concentration  of  the  phosphoric  acid  does      ■ Results
            not   play   a    significant  role  in   the
            determination; it is only necessary to ensure
            that the acid is  not too viscous.  The 85%
            phosphoric acid solution was therefore diluted
            1:5 with water.
            The resulting 17% acid solution was
            transferred into the instrument using an OCT-
            1 autosampler.                                         Abb.: Peak graph of diluted phosphoric acid(17%)

                                                                The 17% phosphoric acid solution measured
                                                                in this way resulted in a TOC concentration of
                                                                0.61  mg/L.  The  relative  standard  deviation
                                                                over three injections was 1.8%.

                                                                ■ Recommended Analyser / Configuration