Page 20 - Food&Beverages Solutions-for-EtO-Analysis-SSI
P. 20

2.2  Chilli Powder Analysis              - Direct Sample Method


                                                                                              Application Databook



                EtO & 2-CE Analysis In Chilli Powder (Direct Sample In Headspace Vial)


                 ¿  As such chilli powder sample preparation into headspace vial for trace level quantitation of EtO
                    & 2-CE. In this approach, chilli powder sample was directly taken into headspace vial and
                    analyzed for its EtO & 2-CE content.

                 ¿  Shimadzu have successfully part validated this method by using Shimadzu’s GCMS-TQ8040 NX
                    with HS-20 NX headspace sampler




                                  APPROACH          MODE OF ANALYSIS   INJECTOR      INSTRUMENT TECHNIQUE

               As such sample in headspace vial       Dynamic Headspace   SPL   HS-20 NX - GCMS-TQ8040 NX
                                (EtO & 2-CE)



                System Description

                   GCMS System                    GCMS-TQ8040 NX
                   Headspace Sampler              HS-20 NX (Dynamic Headspace)
                   Capillary Column               SH-RTX-VMS (60 m X 0.25 mm ID x 1.4 um df)
                   Ionization Mode                Electron Ionization (EI)
                   Sample Information             EtO & 2-CE in chilli powder
                   Sample Preparation             100 % (As such sample in HS vial) + 1 uL of Acetonitrile (Diluent)
                   Spiked Sample Preparation      100 mg chilli powder + 1 uL of 1 ppm standard
                                                  100 mg chilli powder + 1 uL of 2 ppm standard
                                                  100 mg chilli powder + 1 uL of 3 ppm standard
                                                  100 mg chilli powder + 1 uL of 4 ppm standard
                                                  100 mg chilli powder + 1 uL of 5 ppm standard




                Benefits
                 Ÿ  As such sample preparation in headspace vial (without solvent extraction / without any conversion
                   or derivatization  process)
                 Ÿ  Very less sample preparation time
                 Ÿ  Highly cost effective
                 Ÿ  Shimadzu’s optimized solution of dynamic headspace sampler for the analysis of EtO & 2-CE
                   outperforming the current regulatory limits



















                                                           20