Page 23 - Food&Beverages Solutions-for-EtO-Analysis-SSI
P. 23
2.3 & 2.4 Other Spices - Dynamic Headspace Method
Application Databook
EtO & 2-CE Analysis In Other Spices (Extracted solution In Headspace Vial)
2.3 Chicken Masala
APPROACH MODE OF ANALYSIS INJECTOR INSTRUMENT TECHNIQUE
st
Headspace Method 1 (EtO & 2-CE) Dynamic Headspace SPL HS-20 NX - GCMS-TQ8050 NX
2.4 Kitchen Masala
APPROACH MODE OF ANALYSIS INJECTOR INSTRUMENT TECHNIQUE
st
Headspace Method 1 (EtO & 2-CE) Dynamic Headspace SPL HS-20 NX - GCMS-TQ8050 NX
System Description
GCMS System GCMS-TQ8050 NX
Headspace Sampler HS-20 NX (Dynamic Headspace)
Capillary Column RTX-VMS (60 m X 0.45 mm ID x 2.55 um df)
Ionization Mode Electron Ionization (EI)
Sample Information EtO & 2-CE in Other Spices
Reference Method EURL-SRM Analytical Observation report
* Chromatographic conditions & sample preparation method were same as mentioned in application note.
Benefits
Ÿ Shimadzu have developed dynamic headspace methods for the analysis of EtO & 2-CE in spices
matrix. This method outperforms the current regulatory limits.
Ÿ Shimadzus GCMS-TQ8040 NX / GCMS-TQ8050 NX with liquid autosampler & dynamic headspace
sampler is complete tool for the analysis of EtO & 2-CE.
How to reduce acetaldehyde
§ Spices matrix have very high amount of acetaldehyde, and which may interfere with EtO in analysis.
§ So, to reduce acetaldehyde concentration from extracted sample solution, PSA is used as clean up
reagent
§ Amount (25 mg 100 mg ) optimization is require as per the type of matrix
23