Page 7 - Pharmaceutical- Guide to Biopharmaceutical
P. 7

Cell Line Optimization


 Quantitation of Nucleic Acids  BioSpec-nano





 Quantitation of Double-Stranded    benefits

 DNA Using BioSpec-nano                                                                                            Cell Line Optimization
 click here            •  Measure the concentration or check the purity of double-stranded DNA extracts.
                       •  Measure sample quantities as small as 1 μL.
 Operating Principle and Features  Given the steps involved in one set, including measuring double-  •  Automatic wiping function enables a low-carryover system.
 stranded DNA → wiping → adding TE buffer → wiping, repeating that
 The BioSpec-nano has two available optical path lengths, 0.2 mm and     set 60 times resulted in carryover (%) that remained 0.3 % or less,
 0.7 mm, which enable quantitation of nucleic acids in very low sample   which confirmed that sample carryover in the sample area when using
 volumes of 1 or 2 μL. Samples can also be measured using an optional                                              Culture
 cell with a 5 mm optical path length (for 2 mL volumes of dilute   automatic wiping is extremely low.  With the automatic wiping function, never
 samples).   Table 1   Analytical Conditions  forget to wipe off samples.
 An automatic wiping function enables wiping the samples between   Sample concentration  50 to 3700 ng/µL
 measurements, eliminating the need to manually clean the sample stage   Pathlength 0.2 mm  Sample volume  1 µL
 and reducing cross contamination between samples.  Sample concentration  15 to 1000 ng/µL
 Pathlength 0.7 mm
 Sample volume  2 µL
 Measurement Method  5 mm Pathlength Cell   Sample concentration  2 to 150 ng/µL
 (option)  Sample volume  2 mL
 The sample consisted of purified dsDNA dissolved in Tris-EDTA (TE) buffer
 solution. The individual samples were prepared in the concentration ranges
 listed in Table 1 for each pathlength. Next, 10 successive measurements                                           Purification
 were conducted using each of the pathlengths and concentrations using
 the BioSpec-nano, and the OD (Optical Density, absorbance corresponding
 to the 10 mm pathlength) at 260 nm was determined. The Y-axis values
 (Measured OD260) in Fig. 1, 2, and 3 correspond to BioSpec-nano
 measurement values. The standard value (Corrected OD260, X-axis in each
 figure) for determining the accuracy was obtained using the Shimadzu
 Ultraviolet-Visible spectrophotometer, an appropriately diluted sample
 and a 1 mm pathlength cell. The linearities of Fig. 1, 2 and 3 indicate   Fig. 1   Analysis Results with 0.2 mm Pathlength
 the linearity of the standard values, and the deviation from each of the
 straight lines correspond to OD error.
 Results                                                                                                           Characterization

 Analysis Results with 0.2 mm Pathlength  Specifications
 The correlation coefficient of 0.999 for OD260 was obtained with
 respect to the standard value (Fig. 1). When the OD value was greater   Instrument  BioSpec-nano
 than 5 (250 ng/μL dsDNA), the measurement repeatability as CV (%)
 was less than 1.4 %, and the OD error (%) was from -5.4 % to 2.8 %.   Wavelength range  220 to 800 nm
 The data are shown in Fig. 1.   Fig. 2   Analysis Results with 0.7 mm Pathlength  Spectrum bandwidth  3 nm
 Analysis Results with 0.7 mm Pathlength  Wavelength accuracy  ±1 nm                                               Quality Control
 The correlation coefficient of 0.999 for OD260 was obtained with   Pathlength  0.2 mm, 0.7 mm
 respect to the standard value (Fig. 2). When the OD value was greater   Photometric value unit  OD (Optical Density), absorbance converted with 10 mm pathlength
 than 1.4 (70 ng/μL dsDNA), the measurement repeatability as CV (%)
 was less than 1.4 %, and the OD error (%) was from -8.6 % to 4.4 %.   Sample volume  1 µL min. (pathlength: 0.2 mm)
 The data are shown in Fig. 2.        2 µL min. (pathlength: 0.7 mm)
               Light source           Xenon flash lamp
 Analysis Results with 5 mm Pathlength Cell
 The correlation coefficient of 0.999 for OD260 was obtained with   Monochromator  Holographic grating
 respect to the standard value (Fig. 3). When the OD value was greater   Fig. 3   Analysis Results with Optional 5 mm Pathlength Cell  Detector  Photo diode array
 than 0.2 (70 ng/μL dsDNA), the measurement repeatability as CV (%)
 was less than 0.6 %, and the OD error (%) was from -1.6 % to 3.6 %.   Summary  Auto wiping function  Provided     Pharmacokinetics
 The data are shown in Fig. 3.     Spectrum measuring time  3 sec
 BioSpec-nano is capable of simple and excellent measurement linearity,
 Performance of Automatic Wiping in Nucleic Acid Quantitation  reproducibility, and accuracy with a sample volume of 1 to 2 μL for   Quantitation range  Pathlength 0.2 mm, 1 to 75 OD, 50 to 3,700 ng/µL
 We alternated measurement of purified dsDNA (11.7 OD, 578 ng/  optical pathlengths of 0.2 mm and 0.7 mm, respectively.  Pathlength 0.7 mm, 0.3 to 21 OD, 15 to 1,000 ng/µL
                                      Optional 5 mm pathlength cell, 0.04 to 3 OD
 μL) and TE buffer solution using a 0.7 mm pathlength, 3 μL sample
 volume, and 1 wipe operation between measurements. Carryover   Application Examples    Dimensions  W 210 mm × D 214 mm × H 417 mm
 (%) of dsDNA to the TE buffer solution was used as an index of the
 automatic wiping performance.  • Measuring single-strand DNA concentration  Weight  7 kg                          Others
 • Measuring RNA concentrations  Analysis mode  Simple nucleic acid quantitation, labeled nucleic acid quantitation, protein quantitation, labeled
 Carryover (%)   • Measuring protein concentration (refer to p. 34)  protein quantitation, photometric measurement
 [(Nucleic acid concentration in TE measurement)]  Note: The droplet formation status will affect analysis results. Measure quantities that are large enough to enable proper droplet formation.
 = 100×   … (1)
 [(Nucleic acid concentration in dsDNA measurement)]

 6                                                                                                                 7
 index                                                                                                   index
   2   3   4   5   6   7   8   9   10   11   12