Nexera UC UHPLC/SFC

            Efficient Method Development using the UHPLC/SFC Switching System



            When considering separation conditions, using both UHPLC and SFC can help to optimize conditions further.                 The following describes an example of using the UHPLC/SFC switching system to increase the speed of method scouting. The
            The Nexera UC/s UHPLC/SFC switching system provides the ability to use both UHPLC and SFC analysis modes in               Nexera UC/s UHPLC/SFC switching system was used to automatically optimize the separation conditions for two chiral compounds
                                                                                                                                      (omeprazole and warfarin).
            a single system. The  gure below shows a  ow diagram for this system. The system was con gured by adding
                                                                                                                                        A total of 36 combinations of six chiral columns (CHIRALPAK  series) and the three mobile phases shown in the tables below
                                                                                                                                                                                       ®
            a supercritical carbon dioxide delivery unit and back pressure regulator unit to a standard UHPLC system. Both
                                                                                                                                      (18 combinations each for UHPLC and SFC) were evaluated.
            UHPLC and SFC analysis modes can be used by switching of delivery units (control mode ON or OFF) and
                                                                                                                                                                                              Table 1  UHPLC Analytical Conditions for Chiral Compounds
            switching the pressure of the back pressure regulator.                                                                                  SFC                   UHPLC
                                                                                                                                                                                                 Mobile phase
               Sharing the solvent delivery unit (for pumping organic solvents), autosampler, column oven, and detector                    Modi er               Mobile Phase             No.  (Upper: A and Lower: B)   Others
            for both SFC and UHPLC analysis minimizes space requirements and equipment cost and reduces equipment                            No.1   No.2    No.3   No.1    No.2   No.3            Hexane      B Conc. (%)   : 20% (Isocratic)
                                                                                                                                      Column    (5)    (6)            (1)                  1      Ethanol     Flow Rate    : 2 mL/min
            downtime. In addition, an existing UHPLC system can be upgraded to this system.                                              IA                                                                   Column Temperature : 40 °C
                                                                                                                                                                                                              Inj. Vol.    : 1  L
               By using the mobile phase solvent                                                                                                                                           2      Hexane      Detection    : PDA@220 nm
                                                                                                                                                                                                Isopropyl alcohol  Step GE
            switching valve in combination with the                                                                                      IB                                                                       0 - 6 min  20%  Analysis
            column switching valve, mobile phase                                                                                                                                           3  Methyl tertiary butyl ether  6 – 8 min  40%  Column washing
                                                                                                                                                (2)    (4)                                        Ethanol        8 – 12 min  20%  Equilibration
            conditions can be changed automatically                                                                                      IC
            and continuously for up to twelve                                                                                                                                                  Table 2  SFC Analytical Conditions for Chiral Compounds
            columns to enable a wide variety of                                                                                          ID                                               No.     Modi er                Others
                                                                                                                                                                                                              Modifier Conc. (%)  : 20% (Isocratic)
            conditions, improving method                                                                                                                                                   1      Methanol    Flow Rate    : 3 mL/min
                                                                                                                                                                                                              Column Temperature : 40 °C
            development ef ciency.                                                                                                       IE                                                                   Inj. Vol.    : 1  L
                                                                                                                                                                                                              BPR Press    : 10 MPa
                                                                                                                                                                                    (3)    2      Ethanol     Detection    : PDA@220 nm
                                                                                                                                         IF                                                                   Step GE
                                                                                                                                                                                                                  0 - 5 min  20%  Analysis
                                                                                                                                                                                              Acetonitrile / Ethanol  5 – 7 min  40%  Column washing
                                                                                                                                                                                           3
                                                                                                                                              Chromatograms of omeprazole under 36 different     = 75 / 25 (v/v)  7 – 10 min  20%  Equilibration
                                                                                                                                                        analytical conditions

                                                                                                                                        Reproducibility for continuous switching
                                                                        UHPLC column
                                                                   Oven-A                                                             Using the switching system, three drug components were analyzed by continuous switching between UHPLC and SFC
                                 SFC                                                                                                  analysis three times. The resulting chromatograms are shown in Fig. 10. The chromatograms show that reliable results
                                               Pump-A       Column
                                     Shutoff valve:  (for CO2)  switching valve         Manifold                                      were obtained, with no effects from switching  ow lines, even if mobile phases and separation characteristics are
                                       Open
                                                                                                                                      signi cantly different.
                                                                                            Back pressure
                                       Liquid CO2                                         regulator control:                                               Switch from              Switch from             Switch from
                                        cylinder              Flow                            ON                                               SFC (1)    SFC to UHPLC  UHPLC (1)  UHPLC to SFC   SFC (2)   SFC to UHPLC  UHPLC (2)
                                                Pump-B  switching valve                Detector
                                                                                                                                            SFC (n = 5 for each)                        UHPLC (n = 5 for each)
                                                         Autosampler                                                                             Propranolol    Indapamide               Propranolol  Indapamide
                                                                                                                                                Ibuprofen
                                         Mobile phase                                                                                                                                                  Ibuprofen
                        One UHPLC     (organic solvent-based)                                                                                                             After switching
                        pump added to                                                                                                                                     from LC (SFC (2))                          After switching
                        an SFC system                                                                                                                                                                                from SFC (UHPLC (2))
                                                  Pump-C                                                                                                                  Before switching                           After switching
                                                                                                                                                                          (SFC (1))                                  from SFC (UHPLC (1))
                                                                   Oven-B                                                                           1.0          2.0   min                  1.0         2.0       min
                                         Mobile phase                    SFC Column
                                     (water-based or buffer solution)                                                                                Chromatograms for three drug components, obtained by switching between UHPLC and SFC analysis modes
                                                      LC
                            Nexera UC/s UHPLC/SFC switching system and  ow line diagram (Equipment in green frame used for SFC)                                                                          Note: CHIRALPAK ®  is a registered trademark of Daicel Corporation.