Page 7 - Shimadzu LCMS-8050
P. 7

UFscanning





                                                                                           UFsensitivity


                                                                        UFswitching
                                                                                                       UF-MRM










                         Heated ESI Probe
                                                                               Heating gas flow  Nebulizer gas flow
                     In order to improve desolvation efficiency, the newly developed heated ESI probe
                     combines a high-temperature gas with the nebulizer spray, assisting in the
                     desolvation of large droplets and facilitating ionization. This development allows   *  *
                     for high-sensitivity analysis of a wide range of target compounds.



                              100
                                               289.05 > 97.10
                                                                            *The transmission of
                                               Heating gas ON                heat to the LC line
                                                                             has been minimized
                                               Signal intensity:             by mounting the
                               %
                                               30 X improvement              heaters in the
                                                                             bottom-most
                                                                             position.
                                               Heating gas OFF
                                0
                                0.50     1.00    1.50     min
                                    MRM Chromatogram of Testosterone






                         Excellent Reproducibility Even at Attogram (ag) Levels

                     Both sensitivity and reproducibility are essential when establishing low limits of
                     quantitation. High-precision quantitative results obtained with the LCMS-8050 in   100  455.4> 165.0
                                                                                 500 ag in plasma
                     the analysis of Verapamil in blood plasma at levels between 500 ag and 50 pg are
                     shown below. Excellent reproducibility with a % RSD of 2.77 % was obtained when
                     analyzing just 500 ag of Verapamil. The LCMS-8050 demonstrates optimal
                     performance for quantitative analysis of even trace components of a complex matrix.



                                                                             %
                        Concentration  Calculated  % RSD   Accuracy
                          actual    concentration           (%)
                          ng/mL       ng/mL      (n = 6)    (n = 6)
                         0.000500    0.000501     2.77      100.2
                          0.00500     0.00496     3.98      99.2
                           0.0500     0.0506      1.21      101.2
                            0.500      0.493      1.31      98.6              0
                            5.00        4.89      1.81      97.8                  0.8   0.9   1.0    1.1   1.2 min
                            50.0        51.6      0.65      103.2                    MRM Chromatogram of Verapamil





                                                                                 Liquid Chromatograph Mass Spectrometer  /&06       7
   2   3   4   5   6   7   8   9   10   11   12